Diagnostic Screens<br />

Continued<br />

PLEASE WAIT!<br />

...‘ROM’PASSEDTEST<br />

If the ROM fails the diagnostic<br />

test, the sampler reports<br />

’ROM’ FAILED TEST. Contact<br />

<strong>Isco</strong> Customer Service for<br />

assistance.<br />

ABCDEFGHIJKLMNOPQRST<br />

ABCDEFGHIJKLMNOPQRST<br />

ABCDEFGHIJKLMNOPQRST<br />

ABCDEFGHIJKLMNOPQRST<br />

TEST PUMP?<br />

YES NO<br />

PUMPING...<br />

PUMPING...<br />

ON/OFF RATIO = _.__<br />

PURGING...<br />

PURGING...<br />

ON/OFF RATIO = _.__<br />

TEST DISTRIBUTOR<br />

YES NO<br />

FINDING ARM FLEXURE<br />

NOW AT BOTTLE __<br />

RE-INITIALIZE?<br />

YES NO<br />

<strong>6700</strong> SAMPLERS<br />

Distributor Test<br />

The distributor test is provided for factory personnel to verify the<br />

distributor’s position as it rotates through the 24 positions.<br />

(screens 12 through 14). The distributor arm flexes slightly when<br />

it contacts the stop, and the sampler measures this flexure. The<br />

sampler uses this measurement to position the arm accurately<br />

over each bottle.<br />

Reinitialize Controller<br />

The sampler allows you to reset all program settings to the settings<br />

shipped with the sampler (screen 15). Reinitializing also<br />

clears the memory that stores the sampler’s event log and module<br />

data. Use this option cautiously.<br />

MAINTENANCE CHECKLIST<br />

Inspect the pump tube for wear. Replace it if necessary.<br />

Clean the pump tubing housing.<br />

Change the suction line if necessary.<br />

Clean the bottles, suction line, strainer and pump tube.<br />

Check the humidity indicator.<br />

When the battery warning appears on the display, replace the<br />

controller’s internal battery.<br />

CLEANING GUIDELINES<br />

Controller, Top Cover, Center Section, Retaining Ring, and Tub<br />

Clean the top cover, center section, retaining ring, and tub with<br />

warm soapy water or by spraying them with a hose. Avoid using<br />

a high-pressure hose to clean the controller, especially around<br />

the control panel. Extreme pressures may force water past the<br />

control-panel seal.<br />

When cleaning the controller and top cover, cap the connectors at<br />

the back of the controller tightly. Keep a power source connected<br />

to the power-source connector, or replace the protective cap<br />

(shipped with the sampler) over the power source connector, to<br />

protect the pins from moisture damage. Make sure the pump<br />

drain hole (located on the bottom right-hand side of the pump,<br />

beneath the pump band) is open and free of debris or buildup.<br />

Sample Bottles<br />

The sample bottles have a wide mouth to facilitate cleaning.<br />

Wash them with a brush and soapy water, or use a dishwasher.<br />

You can autoclave the glass bottles.<br />

Suction Line, Pump Tube, and Discharge Tube<br />

Clean the suction line, pump tube, and discharge tube by placing<br />

the end of the suction line in a cleaning solution and pumping it<br />

through the delivery system. Rinse with clean water. If these<br />

items are severely contaminated, discard them. For critical<br />

sampling applications, refer to Cleaning Protocols for Priority<br />

Pollutants.<br />

Strainer<br />

Clean the strainer with a brush and soapy water.<br />