Diagnostic Screens<br />
Continued<br />
PLEASE WAIT!<br />
...‘ROM’PASSEDTEST<br />
If the ROM fails the diagnostic<br />
test, the sampler reports<br />
’ROM’ FAILED TEST. Contact<br />
<strong>Isco</strong> Customer Service for<br />
assistance.<br />
ABCDEFGHIJKLMNOPQRST<br />
ABCDEFGHIJKLMNOPQRST<br />
ABCDEFGHIJKLMNOPQRST<br />
ABCDEFGHIJKLMNOPQRST<br />
TEST PUMP?<br />
YES NO<br />
PUMPING...<br />
PUMPING...<br />
ON/OFF RATIO = _.__<br />
PURGING...<br />
PURGING...<br />
ON/OFF RATIO = _.__<br />
TEST DISTRIBUTOR<br />
YES NO<br />
FINDING ARM FLEXURE<br />
NOW AT BOTTLE __<br />
RE-INITIALIZE?<br />
YES NO<br />
<strong>6700</strong> SAMPLERS<br />
Distributor Test<br />
The distributor test is provided for factory personnel to verify the<br />
distributor’s position as it rotates through the 24 positions.<br />
(screens 12 through 14). The distributor arm flexes slightly when<br />
it contacts the stop, and the sampler measures this flexure. The<br />
sampler uses this measurement to position the arm accurately<br />
over each bottle.<br />
Reinitialize Controller<br />
The sampler allows you to reset all program settings to the settings<br />
shipped with the sampler (screen 15). Reinitializing also<br />
clears the memory that stores the sampler’s event log and module<br />
data. Use this option cautiously.<br />
MAINTENANCE CHECKLIST<br />
Inspect the pump tube for wear. Replace it if necessary.<br />
Clean the pump tubing housing.<br />
Change the suction line if necessary.<br />
Clean the bottles, suction line, strainer and pump tube.<br />
Check the humidity indicator.<br />
When the battery warning appears on the display, replace the<br />
controller’s internal battery.<br />
CLEANING GUIDELINES<br />
Controller, Top Cover, Center Section, Retaining Ring, and Tub<br />
Clean the top cover, center section, retaining ring, and tub with<br />
warm soapy water or by spraying them with a hose. Avoid using<br />
a high-pressure hose to clean the controller, especially around<br />
the control panel. Extreme pressures may force water past the<br />
control-panel seal.<br />
When cleaning the controller and top cover, cap the connectors at<br />
the back of the controller tightly. Keep a power source connected<br />
to the power-source connector, or replace the protective cap<br />
(shipped with the sampler) over the power source connector, to<br />
protect the pins from moisture damage. Make sure the pump<br />
drain hole (located on the bottom right-hand side of the pump,<br />
beneath the pump band) is open and free of debris or buildup.<br />
Sample Bottles<br />
The sample bottles have a wide mouth to facilitate cleaning.<br />
Wash them with a brush and soapy water, or use a dishwasher.<br />
You can autoclave the glass bottles.<br />
Suction Line, Pump Tube, and Discharge Tube<br />
Clean the suction line, pump tube, and discharge tube by placing<br />
the end of the suction line in a cleaning solution and pumping it<br />
through the delivery system. Rinse with clean water. If these<br />
items are severely contaminated, discard them. For critical<br />
sampling applications, refer to Cleaning Protocols for Priority<br />
Pollutants.<br />
Strainer<br />
Clean the strainer with a brush and soapy water.<br />
73