Chapter 4 Maintenance<br />
Chapter Contents<br />
Maintenance Screens<br />
SetClock<br />
Diagnostics<br />
Maintenance Checklist<br />
Cleaning Guidelines<br />
Cleaning Protocols for Priority<br />
Pollutants<br />
Replacing the Pump Tube<br />
Checklist for Replacing Pump<br />
Tube<br />
Pump Tube Alarm<br />
Opening the Controller Case<br />
Recharging the Desiccant<br />
Replacing the Internal Battery<br />
Diagnostic Screens<br />
1<br />
2<br />
3<br />
4<br />
RUN<br />
PROGRAM<br />
VIEW REPORT<br />
OTHER FUNCTIONS<br />
Select OTHER FUNCTIONS.<br />
MAINTENANCE<br />
MANUAL FUNCTIONS<br />
PROGRAMMING STYLE<br />
Select MAINTENANCE.<br />
SET CLOCK<br />
SET PUMP TUBE ALARM<br />
INTERNAL BATTERY<br />
DIAGNOSTICS<br />
Select DIAGNOSTICS.<br />
PLEASE WAIT!<br />
. . . ‘RAM’ PASSED TEST<br />
If the RAM fails the diagnostic<br />
test, the sampler reports<br />
’RAM’ FAILED TEST. Contact<br />
<strong>Isco</strong> Customer Service<br />
for assistance.<br />
MAINTENANCE SCREENS<br />
The maintenance screens include:<br />
Set the sampler’s clock.<br />
Set the number of pump counts for the pump tube warning.<br />
See Pump Tube Alarm on page 75.<br />
Reset the battery warning for the internal battery. See Replacing<br />
the Internal Battery on page 79.<br />
Run a set of diagnostic tests on the RAM, ROM (Read Only<br />
Memory), character locations on the display, pump, and distributor,<br />
and to reinitialize the sampler.<br />
Charts for maintenance screens appear in Figure 31, Maintenance<br />
Screens, on page 118.<br />
72<br />
SET CLOCK<br />
<strong>Isco</strong> ships the samplers from the factory with the clocks set to the<br />
correct Central Standard Time. Reset the clock when installing<br />
the sampler in a different time zone, for daylight savings time, or<br />
when needed. Setting the Clock and Calendar on page 19 shows<br />
you how to set the time and date.<br />
DIAGNOSTICS<br />
The sampler has several self-diagnostic tests that check the<br />
memory (both the RAM and ROM), pump, and the distributor<br />
arm position. It also contains screens that let you reset the sampler’s<br />
program settings and memory to factory settings. You will<br />
not need to run the diagnostic tests routinely. The diagnostic<br />
tests are a troubleshooting tool; <strong>Isco</strong>’s Customer Service department<br />
staff may ask you to run these tests when they work with<br />
youindiagnosingproblemswithyoursampler.<br />
Memory Test<br />
Screens 4 and 5 show you the diagnostic tests the memory. If<br />
either screen shows you a message saying that the memory failed<br />
the test, contact <strong>Isco</strong> Customer Service.<br />
Display Test<br />
Screen 6 shows you the display test. You should see the test pattern<br />
illustrated in the screen. If the pattern does not match the<br />
illustration the display or its circuitry require repair.<br />
Pump Test<br />
Screen 7 lets you start the pump test by selecting YES. As the<br />
sampler runs the test, it first runs the pump forward for a short<br />
period. Next, it displays an ON/OFF ratio number. After displaying<br />
the forward pumping ratio, the sampler reverses the pump to<br />
obtain a purge ratio. The acceptable range is between 0.80 and<br />
1.25. If either ratio is outside the acceptable range, the pump<br />
requires repair. You do not need to pump liquid while running the<br />
test, and you can run the test with or without a tube in the pump.