<strong>6700</strong> SAMPLERS<br />

Chapter 4 Maintenance<br />

Chapter Contents<br />

Maintenance Screens<br />

SetClock<br />

Diagnostics<br />

Maintenance Checklist<br />

Cleaning Guidelines<br />

Cleaning Protocols for Priority<br />

Pollutants<br />

Replacing the Pump Tube<br />

Checklist for Replacing Pump<br />

Tube<br />

Pump Tube Alarm<br />

Opening the Controller Case<br />

Recharging the Desiccant<br />

Replacing the Internal Battery<br />

Diagnostic Screens<br />

1<br />

2<br />

3<br />

4<br />

RUN<br />

PROGRAM<br />

VIEW REPORT<br />

OTHER FUNCTIONS<br />

Select OTHER FUNCTIONS.<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

Select MAINTENANCE.<br />

SET CLOCK<br />

SET PUMP TUBE ALARM<br />

INTERNAL BATTERY<br />

DIAGNOSTICS<br />

Select DIAGNOSTICS.<br />

PLEASE WAIT!<br />

. . . ‘RAM’ PASSED TEST<br />

If the RAM fails the diagnostic<br />

test, the sampler reports<br />

’RAM’ FAILED TEST. Contact<br />

<strong>Isco</strong> Customer Service<br />

for assistance.<br />

MAINTENANCE SCREENS<br />

The maintenance screens include:<br />

Set the sampler’s clock.<br />

Set the number of pump counts for the pump tube warning.<br />

See Pump Tube Alarm on page 75.<br />

Reset the battery warning for the internal battery. See Replacing<br />

the Internal Battery on page 79.<br />

Run a set of diagnostic tests on the RAM, ROM (Read Only<br />

Memory), character locations on the display, pump, and distributor,<br />

and to reinitialize the sampler.<br />

Charts for maintenance screens appear in Figure 31, Maintenance<br />

Screens, on page 118.<br />

72<br />

SET CLOCK<br />

<strong>Isco</strong> ships the samplers from the factory with the clocks set to the<br />

correct Central Standard Time. Reset the clock when installing<br />

the sampler in a different time zone, for daylight savings time, or<br />

when needed. Setting the Clock and Calendar on page 19 shows<br />

you how to set the time and date.<br />

DIAGNOSTICS<br />

The sampler has several self-diagnostic tests that check the<br />

memory (both the RAM and ROM), pump, and the distributor<br />

arm position. It also contains screens that let you reset the sampler’s<br />

program settings and memory to factory settings. You will<br />

not need to run the diagnostic tests routinely. The diagnostic<br />

tests are a troubleshooting tool; <strong>Isco</strong>’s Customer Service department<br />

staff may ask you to run these tests when they work with<br />

youindiagnosingproblemswithyoursampler.<br />

Memory Test<br />

Screens 4 and 5 show you the diagnostic tests the memory. If<br />

either screen shows you a message saying that the memory failed<br />

the test, contact <strong>Isco</strong> Customer Service.<br />

Display Test<br />

Screen 6 shows you the display test. You should see the test pattern<br />

illustrated in the screen. If the pattern does not match the<br />

illustration the display or its circuitry require repair.<br />

Pump Test<br />

Screen 7 lets you start the pump test by selecting YES. As the<br />

sampler runs the test, it first runs the pump forward for a short<br />

period. Next, it displays an ON/OFF ratio number. After displaying<br />

the forward pumping ratio, the sampler reverses the pump to<br />

obtain a purge ratio. The acceptable range is between 0.80 and<br />

1.25. If either ratio is outside the acceptable range, the pump<br />

requires repair. You do not need to pump liquid while running the<br />

test, and you can run the test with or without a tube in the pump.