6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>6700</strong> SAMPLERS<br />
COMBINED RESULTS REPORT<br />
At the time of each sample event, the sampler records the readings<br />
from each connected module, rain gauge, or YSI sonde. The<br />
Combined Results Report presents the event time, sample and<br />
bottle number, and the respective readings from the device.<br />
For flow modules, the readings are level and flow rate. For the pH<br />
module, the readings are temperature and pH. For the 4-20 mA<br />
Input module, the readings vary according to your programmed<br />
selections. For the rain gauge and YSI Sonde, the readings vary<br />
according to the selections enabled during the hardware setup.<br />
When recording four or more YSI parameters, the Combined<br />
Results report will exceed the standard 40 character width. To<br />
read the additional columns, the data must be collected with<br />
Flowlink 4 for Windows software.<br />
Combined Results reports appears in Figure 14.<br />
SUMMARY REPORT<br />
This report summarizes the rain gauge or module data. The<br />
typesofreadingsarethesameasthoseavailabletotheCombined<br />
reports. The sampler compiles the data to present a daily<br />
total, the day’s average, maximum, and minimum reading, and a<br />
listing of hourly averages. The sampler then graphs the day’s<br />
readings. Summary reports appear in Figure 15.<br />
Table 10 Reports: Source Codes, Error Codes, and Program Events<br />
SOURCE CODES EXPLANATION<br />
C CALIBRATE SAMPLE Sample volume delivered during calibration.<br />
D DISABLE Sample event at disable time.<br />
E ENABLE Sample event at enable time.<br />
EV EVENT Event-paced sample event.<br />
F FLOW Flow-paced sample event.<br />
G GRABSAMPLE Sample volume delivered as a grab sample.<br />
R RESUME Sample event at resume time.<br />
S START Sample event at start time.<br />
SW SWITCH TIME Sample event at switch time.<br />
TO TIME OVERRIDE Sample event at bottle-switch time.<br />
T TIME Time-paced sample event.<br />
ERROR CODES<br />
DJ DISTRIBUTOR JAMMED The distributor jammed.<br />
IP SAMPLE IN PROGRESS Sample in progress when report was printing.<br />
L PUMP LATCH OPEN Unable to take sample because the pump band was opened.<br />
ND NO DISTRIBUTOR ARM There is no distributor arm or the stop is damaged.<br />
NL NO LIQUID DETECTED The sampler was unable to detect liquid.<br />
NM NO MORE LIQUID After the sampler detected liquid and while the sample was being taken, the liquid detector<br />
stopped detecting liquid.<br />
M BOTTLE FULL The composite bottle is full.<br />
O SAMPLER SHUT OFF The sampler was turned off with the On-Off key during the sample event.<br />
62