<strong>6700</strong> SAMPLERS<br />

COMBINED RESULTS REPORT<br />

At the time of each sample event, the sampler records the readings<br />

from each connected module, rain gauge, or YSI sonde. The<br />

Combined Results Report presents the event time, sample and<br />

bottle number, and the respective readings from the device.<br />

For flow modules, the readings are level and flow rate. For the pH<br />

module, the readings are temperature and pH. For the 4-20 mA<br />

Input module, the readings vary according to your programmed<br />

selections. For the rain gauge and YSI Sonde, the readings vary<br />

according to the selections enabled during the hardware setup.<br />

When recording four or more YSI parameters, the Combined<br />

Results report will exceed the standard 40 character width. To<br />

read the additional columns, the data must be collected with<br />

Flowlink 4 for Windows software.<br />

Combined Results reports appears in Figure 14.<br />

SUMMARY REPORT<br />

This report summarizes the rain gauge or module data. The<br />

typesofreadingsarethesameasthoseavailabletotheCombined<br />

reports. The sampler compiles the data to present a daily<br />

total, the day’s average, maximum, and minimum reading, and a<br />

listing of hourly averages. The sampler then graphs the day’s<br />

readings. Summary reports appear in Figure 15.<br />

Table 10 Reports: Source Codes, Error Codes, and Program Events<br />

SOURCE CODES EXPLANATION<br />

C CALIBRATE SAMPLE Sample volume delivered during calibration.<br />

D DISABLE Sample event at disable time.<br />

E ENABLE Sample event at enable time.<br />

EV EVENT Event-paced sample event.<br />

F FLOW Flow-paced sample event.<br />

G GRABSAMPLE Sample volume delivered as a grab sample.<br />

R RESUME Sample event at resume time.<br />

S START Sample event at start time.<br />

SW SWITCH TIME Sample event at switch time.<br />

TO TIME OVERRIDE Sample event at bottle-switch time.<br />

T TIME Time-paced sample event.<br />

ERROR CODES<br />

DJ DISTRIBUTOR JAMMED The distributor jammed.<br />

IP SAMPLE IN PROGRESS Sample in progress when report was printing.<br />

L PUMP LATCH OPEN Unable to take sample because the pump band was opened.<br />

ND NO DISTRIBUTOR ARM There is no distributor arm or the stop is damaged.<br />

NL NO LIQUID DETECTED The sampler was unable to detect liquid.<br />

NM NO MORE LIQUID After the sampler detected liquid and while the sample was being taken, the liquid detector<br />

stopped detecting liquid.<br />

M BOTTLE FULL The composite bottle is full.<br />

O SAMPLER SHUT OFF The sampler was turned off with the On-Off key during the sample event.<br />

62