6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>6700</strong> SAMPLERS<br />
Grab Sample<br />
When selected, you will be asked to enter the volume to be<br />
pumped. A sample will then be taken as if it is to be placed into a<br />
container outside the sampler base. Do this by removing the<br />
pump tubing from the bulkhead fitting. See Grab Samples on<br />
page 70. A GRAB SAMPLE event will be logged.<br />
Pump Tube Alarm<br />
If the pump tube is changed, this option should be selected to<br />
reset the pump counts on the counter. This will remove the<br />
“WARNING: REPLACE PUMP TUBING” message that may<br />
appear while the program is running. A PUMP TUBE<br />
REPLACED message is logged. See Replacing the Pump Tube on<br />
page 75.<br />
Calibrate Volume<br />
When selected, you will be asked to enter the volume to be<br />
pumped. A sample will then be taken as if it is to be placed into a<br />
container outside the sampler base. Do this by removing the pump<br />
tubing from the bulkhead fitting. See Calibrating on page 71.<br />
After the sample has been pumped, you will then be asked to<br />
enter the amount actually delivered. Once you have entered the<br />
delivered volume and accepted the calibration, a VOLUME<br />
DELIVERED event will be logged.<br />
Cal/Adj Parameters<br />
When the sampler is configured for operation with a module or YSI<br />
sonde, this option becomes available. Selecting this option will display<br />
the appropriate level adjustment screens and/or the calibration<br />
screens. If the level is adjusted, an ADJUSTED LEVEL event is<br />
logged. If a parameter is calibrated, an appropriate event is logged.<br />
Calibrating a parameter probe will temporarily “turn off” the<br />
partition data storage and the sample enable/disable functions.<br />
These functions are disabled during the calibration and for five<br />
minutes after the program is resumed. During this time, parameter<br />
data normally collected at the data storage interval will be<br />
logged as a 252 error message.<br />
Adjust Pacing<br />
This option is available when the running program is paced by<br />
Uniform Time, Flow Pulses, or Flow Volume. If you select this<br />
option, you are asked to enter a new pacing interval. The sampler<br />
logs an INTERVAL CHANGED event. If the sampler is running<br />
a two-part program and both parts are paced by the pacing<br />
listed above, you will be asked to modify both intervals.<br />
The original pacing interval continues to count down until you<br />
return the sampler to the running program (select RESUME<br />
PROGRAM). If the new pacing interval is less than the original<br />
count remaining, the new interval is used.<br />
If the original count remaining is less than the new pacing interval,<br />
the original count will continue to count down to the next sample<br />
event. Subsequent samples are then paced by the new interval.<br />
56