22.03.2013 Views

6700 Portable Sampler User Manual - Isco

6700 Portable Sampler User Manual - Isco

6700 Portable Sampler User Manual - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>6700</strong> SAMPLERS<br />

Grab Sample<br />

When selected, you will be asked to enter the volume to be<br />

pumped. A sample will then be taken as if it is to be placed into a<br />

container outside the sampler base. Do this by removing the<br />

pump tubing from the bulkhead fitting. See Grab Samples on<br />

page 70. A GRAB SAMPLE event will be logged.<br />

Pump Tube Alarm<br />

If the pump tube is changed, this option should be selected to<br />

reset the pump counts on the counter. This will remove the<br />

“WARNING: REPLACE PUMP TUBING” message that may<br />

appear while the program is running. A PUMP TUBE<br />

REPLACED message is logged. See Replacing the Pump Tube on<br />

page 75.<br />

Calibrate Volume<br />

When selected, you will be asked to enter the volume to be<br />

pumped. A sample will then be taken as if it is to be placed into a<br />

container outside the sampler base. Do this by removing the pump<br />

tubing from the bulkhead fitting. See Calibrating on page 71.<br />

After the sample has been pumped, you will then be asked to<br />

enter the amount actually delivered. Once you have entered the<br />

delivered volume and accepted the calibration, a VOLUME<br />

DELIVERED event will be logged.<br />

Cal/Adj Parameters<br />

When the sampler is configured for operation with a module or YSI<br />

sonde, this option becomes available. Selecting this option will display<br />

the appropriate level adjustment screens and/or the calibration<br />

screens. If the level is adjusted, an ADJUSTED LEVEL event is<br />

logged. If a parameter is calibrated, an appropriate event is logged.<br />

Calibrating a parameter probe will temporarily “turn off” the<br />

partition data storage and the sample enable/disable functions.<br />

These functions are disabled during the calibration and for five<br />

minutes after the program is resumed. During this time, parameter<br />

data normally collected at the data storage interval will be<br />

logged as a 252 error message.<br />

Adjust Pacing<br />

This option is available when the running program is paced by<br />

Uniform Time, Flow Pulses, or Flow Volume. If you select this<br />

option, you are asked to enter a new pacing interval. The sampler<br />

logs an INTERVAL CHANGED event. If the sampler is running<br />

a two-part program and both parts are paced by the pacing<br />

listed above, you will be asked to modify both intervals.<br />

The original pacing interval continues to count down until you<br />

return the sampler to the running program (select RESUME<br />

PROGRAM). If the new pacing interval is less than the original<br />

count remaining, the new interval is used.<br />

If the original count remaining is less than the new pacing interval,<br />

the original count will continue to count down to the next sample<br />

event. Subsequent samples are then paced by the new interval.<br />

56

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!