j<br />

k<br />

i<br />

BOTTLE 2<br />

IN 00:12:00<br />

❋❋❋❋❋ gpm 5.73 ft<br />

00001457600 gal<br />

Error Messages<br />

ERRORS HAVE OCCURED<br />

DURING PROGRAM<br />

FATAL ERROR<br />

NO DISTRIBUTOR ARM<br />

<strong>6700</strong> SAMPLERS<br />

If an entire reading is filled with asterisks, the value exceeds the<br />

number of characters that can be displayed. The sampler stores<br />

the actual reading in memory, so it may be collected later.<br />

For those programs that have delayed or scheduled start times,<br />

the readings will be displayed while waiting for the start time.<br />

Keep in mind that any totalizer values will be reset at the start<br />

time. Module and YSI readings are also displayed after a program<br />

is complete, until a key is pressed.<br />

It is not necessary to run a program to view module or YSI readings.<br />

Real time data can be viewed on the display by selecting<br />

“VIEW REPORT” from the main menu. See Viewing the Data on<br />

page 59.<br />

ERROR MESSAGES<br />

If the sampler detects an error that prevents it from taking a<br />

sample or continuing the program, it displays an error message.<br />

There are messages for the following errors:<br />

PUMP JAMMED<br />

DISTRIBUTOR JAMMED<br />

PROBABLE OVERFLOW<br />

POWER FAILED<br />

LOW BATTERY<br />

USER STOPPED<br />

NO DISTRIBUTOR ARM<br />

Since the sampler may encounter more than one error during a<br />

program, it does not display the messages continuously. Instead,<br />

it records each error and the time it occurred in the Sampling<br />

Results and Combined Results report. It alerts you to the<br />

recorded errors by displaying the message, “ERRORS HAVE<br />

OCCURRED DURING PROGRAM.” (See Table 10, Reports:<br />

Source Codes, Error Codes, and Program Events on page 62 for a<br />

description of the errors and events.)<br />

If the sampler detects a low battery condition, it stops the running<br />

program and displays the “LOW BATTERY” error message.<br />

A low battery condition exists when five consecutive sample<br />

events have resulted in a POWER FAILED error message.<br />

Fatal Error<br />

Screen k will only show up if you are using a multiple bottle configuration<br />

and the distributor system fails. Failing the test can<br />

indicate any of the following conditions:<br />

The distributor arm is not attached to the sampler.<br />

The distributor-arm stop inside the center section is missing<br />

or broken.<br />

The distributor drive mechanism inside the controller is malfunctioning.<br />

54<br />

OPERATION ERROR:<br />

PUMP JAMMED<br />

NO LIQUID DETECTED<br />

NO MORE LIQUID<br />

PUMP LATCH OPEN<br />

SAMPLER SHUT OFF<br />

BOTTLE FULL<br />

SAMPLE IN PROGRESS