Error Messages<br />
ERRORS HAVE OCCURED<br />
DURING PROGRAM<br />
FATAL ERROR<br />
NO DISTRIBUTOR ARM<br />
<strong>6700</strong> SAMPLERS<br />
If an entire reading is filled with asterisks, the value exceeds the<br />
number of characters that can be displayed. The sampler stores<br />
the actual reading in memory, so it may be collected later.<br />
For those programs that have delayed or scheduled start times,<br />
the readings will be displayed while waiting for the start time.<br />
Keep in mind that any totalizer values will be reset at the start<br />
time. Module and YSI readings are also displayed after a program<br />
is complete, until a key is pressed.<br />
It is not necessary to run a program to view module or YSI readings.<br />
Real time data can be viewed on the display by selecting<br />
“VIEW REPORT” from the main menu. See Viewing the Data on<br />
page 59.<br />
ERROR MESSAGES<br />
If the sampler detects an error that prevents it from taking a<br />
sample or continuing the program, it displays an error message.<br />
There are messages for the following errors:<br />
PUMP JAMMED<br />
DISTRIBUTOR JAMMED<br />
PROBABLE OVERFLOW<br />
POWER FAILED<br />
LOW BATTERY<br />
USER STOPPED<br />
NO DISTRIBUTOR ARM<br />
Since the sampler may encounter more than one error during a<br />
program, it does not display the messages continuously. Instead,<br />
it records each error and the time it occurred in the Sampling<br />
Results and Combined Results report. It alerts you to the<br />
recorded errors by displaying the message, “ERRORS HAVE<br />
OCCURRED DURING PROGRAM.” (See Table 10, Reports:<br />
Source Codes, Error Codes, and Program Events on page 62 for a<br />
description of the errors and events.)<br />
If the sampler detects a low battery condition, it stops the running<br />
program and displays the “LOW BATTERY” error message.<br />
A low battery condition exists when five consecutive sample<br />
events have resulted in a POWER FAILED error message.<br />
Fatal Error<br />
Screen k will only show up if you are using a multiple bottle configuration<br />
and the distributor system fails. Failing the test can<br />
indicate any of the following conditions:<br />
The distributor arm is not attached to the sampler.<br />
The distributor-arm stop inside the center section is missing<br />
or broken.<br />
The distributor drive mechanism inside the controller is malfunctioning.<br />
54<br />
OPERATION ERROR:<br />
PUMP JAMMED<br />
NO LIQUID DETECTED<br />
NO MORE LIQUID<br />
PUMP LATCH OPEN<br />
SAMPLER SHUT OFF<br />
BOTTLE FULL<br />
SAMPLE IN PROGRESS