Flow Meter<br />

Connector<br />

Rain Gauge<br />

Connector<br />

YSI 600 Sonde<br />

<strong>6700</strong> SAMPLERS<br />

CONNECTING EXTERNAL INSTRUMENTS<br />

Flow Meter Connector<br />

Connect the sampler to <strong>Isco</strong> flow meters, flow loggers, PAL 1101s,<br />

or 1640 Liquid Level Actuators by attaching their connect cable<br />

to the Flow Meter Connector. When using the sampler with an<br />

external instrument, remember these guidelines:<br />

Both the sampler and the flow instrument must be on and<br />

running a program.<br />

Ifusingaflowmeterorflowloggerforflowpacing,thesampler<br />

and the flow instrument must both run programs with<br />

flow pacing settings.<br />

Ifusingaflowmeterorflowloggerfortriggerpacing,thesampler<br />

must run a program with flow pacing settings, and the flow<br />

instrument must run a program with trigger pacing settings.<br />

When the sampler runs an event-paced program, it disregards<br />

pacing pulses from a flow meter or flow logger. However,<br />

the sampler continues to monitor for enable signals.<br />

(See <strong>Sampler</strong> Enable on page 87.)<br />

Rain Gauge Connector<br />

Connect the rain gauge, YSI Sonde, or Refrigerator Temperature<br />

Sensor connect cables to the nine-pin Rain Gauge Connector.<br />

(<strong>Sampler</strong>s with a four-pin Rain Gauge Connector can only accept<br />

a rain gauge.) When connecting a rain gauge and a YSI Sonde, or<br />

rain gauge and a Refrigerator Temperature Sensor, use the<br />

appropriate Y-connect cable. See the Accessories appendix at the<br />

end of this manual.<br />

Notes<br />

Select Extended Programming to use the YSI sonde. (See<br />

Changing to and from Extended and Standard Screens on<br />

page 25.)<br />

Update the Hardware Setup whenever you add or remove a<br />

rain gauge, refrigerator temperature sensor, or YSI sonde.<br />

(See Hardware Setup on page 93.)<br />

In extended programming, rain gauge and YSI sonde readings<br />

are only available if they are selected in the Hardware Setup.<br />

However, when using standard programming, rain gauge readings<br />

are automatically stored in the sampler’s memory.<br />

LOCKING THE SAMPLER<br />

<strong>Isco</strong> offers stainless steel cables to lock the sampler (standard<br />

size – use P/N 68-<strong>6700</strong>-134; compact – use 68-<strong>6700</strong>-111). To lock<br />

the sampler, place the hook through the center section latch.<br />

Repeat this for the other two latches, and then pull the ends of<br />

the cables together over the top of the sampler. Lock the loops<br />

with the padlock. You can also lock each latch individually, using<br />

three padlocks, one for each of the latches.<br />

50