Distributor Arm<br />

Figure 2 Installing the Distributor Arm<br />

Step 1<br />

The two-piece arm has four positions,<br />

marked on the underside<br />

with the letters A, B, C, andD.<br />

Adjust the arm to the correct position<br />

for the bottle kit. Table 6<br />

shows you the correct position for<br />

each kit.<br />

Thread the discharge through the<br />

spring and arm so that the tube<br />

protrudes from the arm 1/16 to 1/8<br />

inch.<br />

Step 2<br />

Attach the arm to the distributor<br />

shaft. The arm and shaft are<br />

keyed so that you can align them<br />

correctly. Secure the arm to the<br />

shaft with the black plastic nut.<br />

Step 3<br />

Slide the end of the discharge<br />

tube onto the bulkhead fitting.<br />

<strong>6700</strong> SAMPLERS<br />

PREPARING THE SAMPLER<br />

INSTALLING THE DISTRIBUTOR ARM<br />

AND DISCHARGE TUBE<br />

When sampling with multiple bottles, both the compact and standard<br />

<strong>6700</strong> <strong>Sampler</strong>s use a distributor arm. It is adjustable so that<br />

you can position the discharge tube over the bottles of each kit.<br />

Each bottle kit also uses a discharge tube of a specific length.<br />

Table 6 liststhecorrectdischargetubelengthsforeachbottlekit.<br />

After installing the distributor arm and discharge tube, run the<br />

distributor to each bottle position to ensure proper installation.<br />

If using bulk tubing to cut your own tubes, cut the ends of the<br />

tube square, not at an angle, and cut the length accurately.<br />

The two-piece distributor arm has four positions, marked on the<br />

undersidewiththelettersA,B,C,andD.Adjustthearmtothe<br />

correct position for the bottle kit being used. Table 6 shows the<br />

correct position for each kit. Thread the discharge tube through<br />

the spring and arm so that the tube protrudes from the arm 1 /16<br />

to 1 /8 inch.<br />

Attach the arm to the distributor shaft. Secure the arm with the<br />

black plastic nut. Slide the free end of the discharge tube onto<br />

the bulkhead fitting.<br />