22.03.2013 Views

6700 Portable Sampler User Manual - Isco

6700 Portable Sampler User Manual - Isco

6700 Portable Sampler User Manual - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>6700</strong> SAMPLERS<br />

Example 6 Storm Water Runoff Sampling (continued)<br />

53<br />

54<br />

55<br />

56<br />

57<br />

58<br />

59<br />

60<br />

61<br />

62<br />

63<br />

64<br />

65<br />

SWITCH BOTTLES ON:<br />

NUMBER OF SAMPLES<br />

TIME<br />

SWITCH BOTTLES EVERY<br />

1 SAMPLES (1- 50)<br />

RUN CONTINUOUSLY?<br />

YES NO<br />

SAMPLE VOLUME:<br />

500 ml (10-1000)<br />

ENABLE:<br />

I/O RAIN LEVEL<br />

FLOW Y-pH Y-TEMP<br />

Y-COND Y-D.O. NONE<br />

ENABLE: RAIN<br />

AND OR DONE<br />

ENABLE: RAIN AND<br />

I/O RAIN LEVEL<br />

FLOW Y-pH Y-TEMP<br />

Y-COND Y-D.O. NONE<br />

“RAIN” SET POINT:<br />

0.15 INCHES PER<br />

15 MIN 30 MIN<br />

1 HOUR 2 HOURS<br />

ENABLED WHEN:<br />

ABOVE SET POINT<br />

BELOW SET POINT<br />

RESET RAIN HISTORY?<br />

YES NO<br />

“LEVEL” CONDITION:<br />

SET POINT<br />

RANGE<br />

“LEVEL” SET POINT<br />

0.25 ft<br />

(0.100-30.000)<br />

ENABLED WHEN:<br />

ABOVE SET POINT<br />

BELOW SET POINT<br />

Select NUMBER OF SAMPLES.<br />

Enter 1.<br />

Select NO.<br />

Enter 500.<br />

Select RAIN.<br />

Select AND.<br />

Select LEVEL.<br />

Enter 0.15 for inches, then select 30 minutes.<br />

Select ABOVE SET POINT.<br />

Select YES.<br />

Select SET POINT.<br />

Enter 0.25.<br />

Select ABOVE SET POINT.<br />