<strong>6700</strong> SAMPLERS<br />

Example 6 Storm Water Runoff Sampling<br />

Storm water runoff sampling is an ideal application for two-part<br />

programming. Typical monitoring permits require that a series<br />

of timed samples be taken during the initial portion of the storm<br />

event, followed by flow paced samples during the remainder of<br />

the event.<br />

Note<br />

To enable the Bubbler Module<br />

programming screens, insert the<br />

module before turning on the<br />

sampler.<br />

This example assumes that a Rain Gauge and a Bubbler Flow<br />

Module are connected to the sampler. Using a 24-bottle kit, it<br />

divides the bottles into two groups, with six bottles assigned to<br />

part ‘A’ and 18 to part ‘B.’ The sampler will begin taking samples<br />

after being enabled according to programmable enable controls;<br />

that is, when the rain gauge detects 0.15 inches of rainfall in 30<br />

minutes, and the module detects a flow stream level of more than<br />

0.25 feet. Because both program parts use the same enable settings,<br />

the sampler will enable both parts at the same time. A list<br />

of settings for the program follow:<br />

Program type: Extended.<br />

Program name: STORM.<br />

Site description: SITE 54.<br />

Units: Length–Ft,FlowRate–gpm,FlowVolume–gal<br />

Mode of Operation: Flowmeter, Flow-Insert, Round, 12"<br />

Bottle kit: 24, 1000-milliliter bottles.<br />

Suction-line length: 5 feet.<br />

Suction head: Auto suction head.<br />

Line rinses: None.<br />

Sampling retries: None.<br />

Part ‘A’ and ‘B’ bottle assignments: Bottles1-6topart‘A’,<br />

bottles 7 - 24 to part ‘B’.<br />

Part ‘A’ pacing: Time pacing, sampling every 5 minutes.<br />

Part ‘A’ distribution: Sequential, 1 sample per bottle.<br />

Part ‘A’ sample volume: 800 milliliters.<br />

Part ‘A’ enable: At least 0.15 inches of rainfall in 30 minutes<br />

and a flow stream level of more than 0.25 feet; once enabled,<br />

stay enabled, sample taken when the sampler is enabled.<br />

Part ‘A’ pause and resumes: None.<br />

Part ‘B’ pacing: Flow pacing, sampling every 2,500 gallons.<br />

Part ‘B’ distribution: Sequential, 1 sample per bottle.<br />

Part ‘B’ sample volume: 500 milliliters.<br />

Part ‘B’ enable: At least 0.15 inches of rainfall in 30 minutes<br />

and a flow stream level of more than 0.25 feet; once enabled,<br />

stay enabled, no sample taken when the sampler is enabled;<br />

first sample taken 20 minutes after the sampler is enabled.<br />

Part ‘B’ pause and resumes: None.<br />

Start time: Delayed start of 1 minute.<br />

730 Bubbler Module: Installed<br />