6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
About Stored Programs<br />
After selecting a stored program,<br />
review the settings with the quick<br />
view screens. Each sampling program<br />
stores all the settings<br />
required by the sampler to run a<br />
program. These settings include<br />
the bottle-kit information (number<br />
of bottles and the bottle volume),<br />
the suction-line length, the site<br />
description, start times, and so on.<br />
Ifyouusethesameprogramat<br />
two sites — each requiring different<br />
bottle configurations, suction<br />
line lengths, or other details —<br />
always check these settings before<br />
running a program. You may need<br />
to make minor changes to the program<br />
to make sure the settings<br />
match your equipment.<br />
Stored Programs and<br />
Reinitializing<br />
Reinitializing the sampler restores<br />
the program settings to the factory<br />
settings shipped with the sampler.<br />
Only reinitialize the sampler<br />
when you can afford to lose the<br />
program settings.<br />
<strong>6700</strong> SAMPLERS<br />
Two-part programming is ideal for sampling storm events. Many<br />
storm-water run-off monitoring permits require a first-flush grab<br />
sample within the first 30 minutes of the storm event and flowpaced<br />
samples during the remainder of the event. With a twopart<br />
program, you can set up part ‘A’ for the first-flush sample<br />
and part ‘B’ for the remainder. Example 6 on page 31 shows a<br />
storm-water sampling program.<br />
STORAGE FOR EXTENDED PROGRAMS<br />
The sampler stores five sampling programs: one standard and<br />
four extended. Program storage eliminates the need to reprogram<br />
the sampler with frequently used settings. This feature is<br />
especially useful for complex extended programs such as two-part<br />
programs or programs with many nonuniform-time settings.<br />
<strong>Isco</strong> ships the sampler to you with factory programs that you can<br />
modify to suit your own needs. All four extended programs are<br />
identical to each other.<br />
The stored programs will also save your 700 Series Module<br />
program settings and any sampler enable conditions using rain<br />
gauge data or YSI sonde parameters. These program settings<br />
remain as long as you do not change the type of module, or<br />
change the rain gauge or YSI Hardware Setup. If you make any<br />
of these changes, the sampler updates the stored programs to the<br />
current sampler configuration.<br />
Selecting a Stored Program<br />
The sampler shows the extended programming screens, it always<br />
has one current and three noncurrent programs, each occupying<br />
a permanent location in memory. The first time you add the<br />
extended screens, the current program is “EXTENDED 1,” one of<br />
the factory extended programs. The name of the current<br />
extended program appears in quotations on the first line of the<br />
main menu (See Example 4 on page 27).<br />
The factory programs are named “EXTENDED 1,” “EXTENDED<br />
2,” “EXTENDED 3,” and “EXTENDED 4." You can rename them<br />
with more descriptive names, making them easier to identify.<br />
The program name is one of the extended program’s settings.<br />
The current program is the program you see when you select<br />
PROGRAM from the start-up menu and page through the<br />
programming screens. It is the program the sampler uses when<br />
you select RUN from the start-up menu.<br />
You cannot delete programs, but you can modify them as often as<br />
necessary. However, to run a program with different settings<br />
without losing the settings in your current program, select a<br />
stored program and modify its settings.<br />
To understand how this works, consider this example. You have<br />
been running the current program, named “EXTENDED 1,” at<br />
site A for several weeks. This program contains all the settings<br />
you need for site A, and you don’t want to lose those settings.<br />
26