6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>6700</strong> SAMPLERS<br />
Table 3: Technical Specifications for the <strong>6700</strong> Compact and Standard <strong>Sampler</strong>s<br />
<strong>6700</strong> CONTROLLER: SOFTWARE SPECIFICATIONS (CONTINUED)<br />
YSI 600 Sonde Readings: pH 8 bits 0.1 pH storage resolution<br />
Temp. 16 bits 0.1° C storage resolution<br />
D.O. 8 bits 0.1 mg/l storage resolution<br />
Conductivity 16 bits representing:<br />
Specific Conductance 0.002 mS/cm storage resolution<br />
T.D.S. 0.002 g/l storage resolution<br />
Salinity 0.1 ppt storage resolution<br />
Conductivity 0.002 mS/cm storage resolution<br />
Module Reading Conversions: Level and 4–20 mA readings converted to flow rate units:<br />
gallons per second gallons per minute<br />
million gallons per day cubic feet per second<br />
cubic meters per second cubic meters per hour<br />
cubic meters per day liters per second<br />
Velocity units:<br />
feet per second meters per second<br />
Totalized flow units:<br />
gallons million gallons<br />
cubic feet cubic meters<br />
liters<br />
Devices supported in flow conversion:<br />
Weirs: V-notch; 22.5, 30, 45, 60, 90, and 120 degrees<br />
Rectangular; with and without end contractions<br />
Cipoletti.<br />
Flumes:Parshall; 1", 2", 3", 6", 9", 1.5', 2', 3', 4', 5', 6', 8', 10', 12'<br />
Palmer-Bowlus; 4", 6", 8", 12", 15", 18", 21", 24", 27", 30", 48"<br />
Trapezoidal; LG 60 V, 2" 45 WSC, 12" 45 SRCRC<br />
H; 0.5', 0.75', 1', 1.5', 2', 2.5', 3', 4.5'<br />
Manning Formula: Round, U-channel, Rectangular, Trapezoidal<br />
Area Velocity: Round, U-channel, Rectangular, Trapezoidal<br />
Data Points: 50 level-flow rate points, 50 level-area points<br />
YSI 600 Sonde Units: pH: 2.0 to 14.0<br />
Temperature:°Cor°F<br />
Dissolved Oxygen: mg/l (milligrams per liter)<br />
Conductivity reported as –<br />
Specific Conductance: mS/cm (millisiemens per centimeter)<br />
Salinity: ppt (parts per thousand)<br />
Total Dissolved Solids: g/l (grams per liter)<br />
Conductivity: mS/cm (millisiemens per centimeter)<br />
Refrigerator Temperature<br />
Sensor<br />
Optional sensor and 5 ft (1.55 m) cable to record temperature readings.<br />
Intended for use with fiberglass refrigerated (FR) samplers.<br />
Range: -40.0 °C to 100 °C<br />
Accuracy specification: ±0.3 °C over a range of 0.0 to 10.0 °C.<br />
Reading data is 2 bytes, with a storage resolution of 0.1 °C.<br />
10