<strong>6700</strong> SAMPLERS<br />
Table 1: <strong>6700</strong> Standard and Compact <strong>Sampler</strong> Features<br />
EXTENDED PROGRAMMING FEATURES (CONTINUED)<br />
Distribution Methods Composite: Samples deposited in a single large bottle.<br />
Sequential: Only one sample placed in each bottle.<br />
Samples Per Bottle: Multiple samples placed in each bottle<br />
Bottles Per Sample: One sample deposited in multiple bottles<br />
Multiple Bottle Compositing: A combination of samples per bottle and<br />
bottles per sample distribution methods.<br />
Time Switched Bottles or Bottle sets: Control the sample distribution<br />
using clock times.<br />
See Distribution on page 84.<br />
Programmable <strong>Sampler</strong><br />
Enable<br />
A <strong>6700</strong> sampler can be programmed to enable or disable a running sampling<br />
program when readings received from a connected rain gauge, module, or YSI<br />
Sonde meet certain conditions.<br />
See <strong>Sampler</strong> Enable on page 87.<br />
Pauses and Resumes Create intermittent sampling schedules.<br />
See Pauses/Resumes on page 89.<br />
Two-Part Programming Two-part programming lets you set up a sampling program that divides the<br />
bottles into two groups, filling each group according to separate pacing,<br />
distribution, sampler enable, and pause and resume settings. Ideal for stormwater<br />
run-off sampling.<br />
See One-Part and Two-Part Programs on page 25.<br />
Auto Suction Head or Fixed<br />
Suction Head<br />
The suction head, is the vertical distance from the flow stream to the liquid<br />
detector. Extended programming has two settings.<br />
Auto-Suction Head: The head is automatically determined.<br />
Fixed Suction Head: A user-definable measurement for the head.<br />
Note: See Suction Head on page 81.<br />
Suction Line Rinses Program setting for the number of times (0 to 3) that the <strong>6700</strong> rinses the<br />
suction line before drawing a sample.<br />
See Rinses and Retries on page 82.<br />
Sampling Retries Program setting for the number of times (0 to 3) that the <strong>6700</strong> attempts to<br />
sample if it fails to deliver the entire sample volume.<br />
See Rinses and Retries on page 82.<br />
Three Flexible<br />
Start Time Settings<br />
Option for a Continuous<br />
Running Program<br />
Run Immediately: Starts the sampling program immediately.<br />
Delayed Start: Starts the sampling program after a user-definable delay of<br />
1to999minutes.<br />
Clock Time: Starts the sampling program at a user-definable time on one<br />
or more days of the week.<br />
See Start Times on page 89.<br />
Continuous Sampling: When sample bottles are regularly replaced, the<br />
sample distribution can restart with the first bottle set after the last bottle<br />
set is filled, without interrupting the running program.<br />
See Distribution on page 84.<br />