<strong>6700</strong> SAMPLERS<br />

The software version number changes when your sampler’s software<br />

is updated. Screens 56 through 60 show you how to find the<br />

system IDs.<br />

101<br />

MEMORY<br />

The <strong>6700</strong> <strong>Sampler</strong> contains 128 kilobytes of battery backed RAM<br />

(Random Access Memory) that stores:<br />

Five sampling programs<br />

The sampling results from the most recent program<br />

The rainfall data collected from rain gauges<br />

The data from a refrigerator temperature sensor<br />

The data collected from a 700 Series module<br />

Parameter data collected from a YSI Sonde<br />

Calculated flow information from the current flow module<br />

The sampling results memory stores up to 1004 events. The program<br />

run and start time information is logged as four events.<br />

These four entries are held for the duration of the program. The<br />

remaining 1000 are for logging sample events, program events, and<br />

errors from the currently running program. If the running program<br />

exceeds 1000 events, the data will “roll-over.” That is, the sampler<br />

will replace the oldest stored event with the newest event. When<br />

using a YSI Sonde the sampler logs two entries at each sample<br />

event. This reduces the sampling results memory to 500 events.<br />

Module, YSI Sonde, Refrigerator Temperaure, and rain gauge<br />

readings are stored in memory partitions. The storage capacity<br />

depends on the reading types, the number of bytes per reading, and<br />

the data storage interval.<br />

The sampler clears the sampling results memory when the site<br />

description changes. It is important to collect the sampling<br />

results before changing the site description – the data can not be<br />

retrieved afterwards. Keep in mind that the description may<br />

change when selecting an extended program.<br />

The partition data is cleared when the sampler automatically<br />

creates and sizes the memory partitions. This occurs when you:<br />

Select a new program<br />

Change the Data Storage Interval in the current program<br />

Change the programming mode<br />

Install a new 700 Series Module<br />

Change 700 Series Module mode of operation<br />

Add or remove a rain gauge with the Hardware Setup<br />

Add or remove a refrigerator temperaure sensor as I/O3<br />

Add or remove a YSI parameter with the Hardware Setup<br />

Reinitialize the sampler