6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
6700 Portable Sampler User Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Analog Output Screens<br />
Charts for analog output screens<br />
appear in:<br />
Figure 47, Analog Output and<br />
Dialout Screens on page 133<br />
Figure 48, Analog Output and<br />
Dialout Quick View Screens on<br />
page 133.<br />
53<br />
54<br />
55<br />
Programming the serial<br />
output<br />
PERIODIC<br />
SERIAL OUTPUT?<br />
YES NO<br />
PORT SETTINGS:<br />
9600N81 4800N81<br />
2400N81 1200N81<br />
SERIAL OUTPUT EVERY:<br />
15 SEC 1 MIN<br />
5MIN 15MIN<br />
<strong>6700</strong> SAMPLERS<br />
Refrigerator Temperature<br />
<strong>Isco</strong> offers a temperature sensor intended for use with the<br />
<strong>6700</strong>FR Refrigerated <strong>Sampler</strong>. The optional sensor connects to<br />
the Rain Gauge/YSI port and uses I/O3 to transmit its temperature<br />
readings to the <strong>6700</strong>.<br />
To configure an attached temperature sensor, select “FR-TEMP”<br />
as the input for I/O3. The <strong>6700</strong> will display the NEW HARD-<br />
WARE SETUP screen. Select YES and the <strong>6700</strong> reconfigures its<br />
memory partitions to record the temperature readings.<br />
Temperature readings are updated approximately every 2.5 seconds.<br />
The readings are stored as FR-TEMP readings at the data<br />
storage interval. You can retrieve a history of temperature readings<br />
through the <strong>6700</strong>’s reports. The temperature data also can<br />
be used as a sampler enable condition.<br />
ANALOG OUTPUT<br />
<strong>Isco</strong> offers multiple analog outputs as a factory installed option.<br />
The number of isolated outputs can vary from one to three and is<br />
specified at order time. The option is installed inside the controller<br />
and includes a six-pin connector to interface with external<br />
devices.<br />
The analog outputs can be programmed for either 0-to-20 mA or<br />
4-to-20 mA operation. With the exception of rain, any parameter<br />
that the <strong>6700</strong> sampler is currently measuring can be used to control<br />
any (or all) of the analog output channels. Programming the<br />
outputs requires parameter levels at each end of the analog<br />
range and then assumes a linear relationship between those<br />
points. MANUAL CONTROL of each output is also available.<br />
SERIAL OUTPUT<br />
Serial ASCII data can be output through the controller’s interrogator<br />
port. The output frequency is user-selectable: every 15 seconds,<br />
every minute, every 5 minutes, every 15 minutes, or by<br />
command.<br />
To receive serial data as a periodic output:<br />
1. Select YES at screen 53.<br />
2. Select the appropriate baud rate setting in screen 54.Notethat<br />
at all baud rates the data is sent with no parity, eight data bits,<br />
and one stop bit. If you have difficulty with the connection, try<br />
a slower baud rate.<br />
3. Select the desired output interval in screen 55.<br />
4. Connect to the <strong>6700</strong> controller’s interrogator connector using a<br />
Serial Output interrogator cable (sense line open, P/N<br />
60-9004-263). Do not use a standard interrogator cable.<br />
To receive serial data by command:<br />
1. Select NO at screen 53.<br />
2. Connect your computer to the <strong>6700</strong> controller’s interrogator<br />
connector using a standard interrogator cable (sense line<br />
grounded), P/N 60-2544-040 (25 pin) or 60-2544-044 (9 pin).<br />
98