Analog Output Screens<br />

Charts for analog output screens<br />

appear in:<br />

Figure 47, Analog Output and<br />

Dialout Screens on page 133<br />

Figure 48, Analog Output and<br />

Dialout Quick View Screens on<br />

page 133.<br />

53<br />

54<br />

55<br />

Programming the serial<br />

output<br />

PERIODIC<br />

SERIAL OUTPUT?<br />

YES NO<br />

PORT SETTINGS:<br />

9600N81 4800N81<br />

2400N81 1200N81<br />

SERIAL OUTPUT EVERY:<br />

15 SEC 1 MIN<br />

5MIN 15MIN<br />

<strong>6700</strong> SAMPLERS<br />

Refrigerator Temperature<br />

<strong>Isco</strong> offers a temperature sensor intended for use with the<br />

<strong>6700</strong>FR Refrigerated <strong>Sampler</strong>. The optional sensor connects to<br />

the Rain Gauge/YSI port and uses I/O3 to transmit its temperature<br />

readings to the <strong>6700</strong>.<br />

To configure an attached temperature sensor, select “FR-TEMP”<br />

as the input for I/O3. The <strong>6700</strong> will display the NEW HARD-<br />

WARE SETUP screen. Select YES and the <strong>6700</strong> reconfigures its<br />

memory partitions to record the temperature readings.<br />

Temperature readings are updated approximately every 2.5 seconds.<br />

The readings are stored as FR-TEMP readings at the data<br />

storage interval. You can retrieve a history of temperature readings<br />

through the <strong>6700</strong>’s reports. The temperature data also can<br />

be used as a sampler enable condition.<br />

ANALOG OUTPUT<br />

<strong>Isco</strong> offers multiple analog outputs as a factory installed option.<br />

The number of isolated outputs can vary from one to three and is<br />

specified at order time. The option is installed inside the controller<br />

and includes a six-pin connector to interface with external<br />

devices.<br />

The analog outputs can be programmed for either 0-to-20 mA or<br />

4-to-20 mA operation. With the exception of rain, any parameter<br />

that the <strong>6700</strong> sampler is currently measuring can be used to control<br />

any (or all) of the analog output channels. Programming the<br />

outputs requires parameter levels at each end of the analog<br />

range and then assumes a linear relationship between those<br />

points. MANUAL CONTROL of each output is also available.<br />

SERIAL OUTPUT<br />

Serial ASCII data can be output through the controller’s interrogator<br />

port. The output frequency is user-selectable: every 15 seconds,<br />

every minute, every 5 minutes, every 15 minutes, or by<br />

command.<br />

To receive serial data as a periodic output:<br />

1. Select YES at screen 53.<br />

2. Select the appropriate baud rate setting in screen 54.Notethat<br />

at all baud rates the data is sent with no parity, eight data bits,<br />

and one stop bit. If you have difficulty with the connection, try<br />

a slower baud rate.<br />

3. Select the desired output interval in screen 55.<br />

4. Connect to the <strong>6700</strong> controller’s interrogator connector using a<br />

Serial Output interrogator cable (sense line open, P/N<br />

60-9004-263). Do not use a standard interrogator cable.<br />

To receive serial data by command:<br />

1. Select NO at screen 53.<br />

2. Connect your computer to the <strong>6700</strong> controller’s interrogator<br />

connector using a standard interrogator cable (sense line<br />

grounded), P/N 60-2544-040 (25 pin) or 60-2544-044 (9 pin).<br />