Figure 23 Event Mark Timing Diagram<br />

50<br />

51<br />

52<br />

Pulse at Beginning of<br />

Presample Purge<br />

Pulse at Beginning of<br />

Forward Pumping<br />

Variable Duration Pulse<br />

During Entire Pump<br />

Variable Duration Pulse<br />

During Forward<br />

Pumping<br />

<strong>6700</strong> SAMPLERS<br />

Presample Purge Rinse Cycle Pump Volume 1 Pump Volume 2 Pump Volume 3 Postsample Purge<br />

Note: This diagram shows the pulses sent for a bottles-per-sample distribution scheme<br />

Using the Purge-Count<br />

screens<br />

PRE-SAMPLE PURGE:<br />

____COUNTS<br />

(10-9999)<br />

POST-SAMPLE PURGE:<br />

DEPENDENT ON HEAD<br />

FIXED COUNT<br />

POST-SAMPLE PURGE:<br />

____COUNTS<br />

(10-9999)<br />

Note<br />

The state of the I/O lines is not<br />

defined when the sampler is OFF.<br />

PRESAMPLE AND POSTSAMPLE<br />

PURGE COUNTS<br />

The presample purge is when the pump runs in reverse clearing<br />

the strainer of any debris before taking a sample. The presample<br />

purge is set at 200 pump counts. After the sample is taken a postsample<br />

purge occurs to clear the liquid from the suction line. The<br />

duration of the postsample purge is determined by the length of<br />

the suction line to maximize the battery life. The pre- and postsample<br />

purge can also be manually set.<br />

I/O PIN PROGRAMMING<br />

Pins C, H, and I on the Rain Gauge Connector are user-configured<br />

as I/O1, I/O2, and I/O3 (Input/Output pins) respectively.<br />

Each pin can be either an input for sampler enabling, or an output<br />

for alarms such as PROGRAM DONE, RUN ERROR, etc.<br />

The CONDITIONS I/O output can report alarm conditions using<br />

the parameters the <strong>6700</strong> is configured to measure. Similar to the<br />

<strong>Sampler</strong> Enable, you can set thresholds for one or two conditions<br />

and the CONDITIONS output can then trigger alarms when the<br />

measured parameters exceed the threshold.<br />

<strong>Isco</strong> offers an I/O module that interfaces the sampler with a large<br />

number of non-<strong>Isco</strong> devices. For more information about the I/O<br />

module, contact your sales representative.<br />

97<br />

Event Mark<br />

Bottle Number<br />

Event Mark<br />

Bottle Number<br />

Event Mark<br />

Bottle Number<br />

Event Mark<br />

Bottle Number