Figure 23 Event Mark Timing Diagram<br />
50<br />
51<br />
52<br />
Pulse at Beginning of<br />
Presample Purge<br />
Pulse at Beginning of<br />
Forward Pumping<br />
Variable Duration Pulse<br />
During Entire Pump<br />
Variable Duration Pulse<br />
During Forward<br />
Pumping<br />
<strong>6700</strong> SAMPLERS<br />
Presample Purge Rinse Cycle Pump Volume 1 Pump Volume 2 Pump Volume 3 Postsample Purge<br />
Note: This diagram shows the pulses sent for a bottles-per-sample distribution scheme<br />
Using the Purge-Count<br />
screens<br />
PRE-SAMPLE PURGE:<br />
____COUNTS<br />
(10-9999)<br />
POST-SAMPLE PURGE:<br />
DEPENDENT ON HEAD<br />
FIXED COUNT<br />
POST-SAMPLE PURGE:<br />
____COUNTS<br />
(10-9999)<br />
Note<br />
The state of the I/O lines is not<br />
defined when the sampler is OFF.<br />
PRESAMPLE AND POSTSAMPLE<br />
PURGE COUNTS<br />
The presample purge is when the pump runs in reverse clearing<br />
the strainer of any debris before taking a sample. The presample<br />
purge is set at 200 pump counts. After the sample is taken a postsample<br />
purge occurs to clear the liquid from the suction line. The<br />
duration of the postsample purge is determined by the length of<br />
the suction line to maximize the battery life. The pre- and postsample<br />
purge can also be manually set.<br />
I/O PIN PROGRAMMING<br />
Pins C, H, and I on the Rain Gauge Connector are user-configured<br />
as I/O1, I/O2, and I/O3 (Input/Output pins) respectively.<br />
Each pin can be either an input for sampler enabling, or an output<br />
for alarms such as PROGRAM DONE, RUN ERROR, etc.<br />
The CONDITIONS I/O output can report alarm conditions using<br />
the parameters the <strong>6700</strong> is configured to measure. Similar to the<br />
<strong>Sampler</strong> Enable, you can set thresholds for one or two conditions<br />
and the CONDITIONS output can then trigger alarms when the<br />
measured parameters exceed the threshold.<br />
<strong>Isco</strong> offers an I/O module that interfaces the sampler with a large<br />
number of non-<strong>Isco</strong> devices. For more information about the I/O<br />
module, contact your sales representative.<br />
97<br />
Event Mark<br />
Bottle Number<br />
Event Mark<br />
Bottle Number<br />
Event Mark<br />
Bottle Number<br />
Event Mark<br />
Bottle Number