Selecting Hardware Setup<br />
41<br />
42<br />
43<br />
44<br />
RUN “PROGRAM NAME”<br />
PROGRAM<br />
VIEW REPORT<br />
OTHER FUNCTIONS<br />
Select Other Functions<br />
MAINTENANCE<br />
MANUAL FUNCTIONS<br />
PROGRAMMING STYLE<br />
HARDWARE<br />
Select Hardware<br />
Liquid Detector Disable<br />
USE LIQUID DETECTOR?<br />
YES NO<br />
Rain Gauge Setup<br />
DO YOU HAVE A<br />
RAIN GAGE CONNECTED?<br />
YES NO<br />
Select YES.<br />
RAIN GAGE:<br />
0.01 inch TIP<br />
0.1 mm TIP<br />
The rain gauge is calibrated<br />
for inches or millimeters.<br />
Select the calibration of your<br />
rain gauge.<br />
<strong>6700</strong> SAMPLERS<br />
The QUICK VIEW/CHANGE programming style includes a series<br />
of “summary” screens. Summary screens include one or more program<br />
settings. You can quickly step through the program by selecting<br />
the ↓ located in the lower right-hand corner of the display. You<br />
canevenbackstepthroughtheprogrambyselectingthe↑ located in<br />
the upper left hand corner of the display. Program changes are made<br />
by selecting the program setting you wish to change. Then make the<br />
changes as you would in the NORMAL programming mode. The<br />
extended programming mode defaults to QUICK VIEW/CHANGE.<br />
93<br />
HARDWARE SETUP<br />
The hardware setup screens, available with extended programming,<br />
make a number of adjustments to the <strong>Sampler</strong>’s hardwarebased<br />
features:<br />
Liquid Detector Disable Rain Gauge Setup<br />
YSI 600 Sonde Setup Master/slave Sampling<br />
Full Bottle Detection Display Backlighting<br />
Event Marks Purge Counts<br />
I/O Pin Programming Analog Outputs<br />
Serial Output Dialout Alarms<br />
LIQUID DETECTOR DISABLE<br />
Under some conditions (pressurized lines, sampling foam), a more<br />
repeatable sample volume may be delivered by disabling the liquid<br />
detector. Selecting NO will turn the Liquid Detector off.<br />
If the Liquid Detector is off:<br />
you must enter the SUCTION HEAD,<br />
no RINSES or RETRIES will be available,<br />
you will not be able to CALIBRATE SAMPLE volumes,<br />
the RESULTS report will read a COUNT TO LIQUID of 0,<br />
a complete purge will occur between samples when using multiple<br />
bottles-per-sample distribution.<br />
RAIN GAUGE SETUP<br />
With standard programming, the sampler automatically records<br />
the rain gauge readings. However, with extended programming,<br />
recording the rain gauge readings and the rain enable option can<br />
be turned on or off. Enter the hardware setup and follow screens<br />
43 and 44 in the margin.<br />
At screen 43, select YES to record the rain gauge readings and<br />
add RAIN as a sampler enable option. Continue with screen 44<br />
and the sampler will then create a rain memory partition and<br />
update the available sampler enable conditions.<br />
Select NO to disable the readings and the sampler enable option.<br />
When you select NO, the sampler removes the rain memory partition,<br />
leaving the memory space available for other reading types.