Selecting Hardware Setup<br />

41<br />

42<br />

43<br />

44<br />

RUN “PROGRAM NAME”<br />

PROGRAM<br />

VIEW REPORT<br />

OTHER FUNCTIONS<br />

Select Other Functions<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

Select Hardware<br />

Liquid Detector Disable<br />

USE LIQUID DETECTOR?<br />

YES NO<br />

Rain Gauge Setup<br />

DO YOU HAVE A<br />

RAIN GAGE CONNECTED?<br />

YES NO<br />

Select YES.<br />

RAIN GAGE:<br />

0.01 inch TIP<br />

0.1 mm TIP<br />

The rain gauge is calibrated<br />

for inches or millimeters.<br />

Select the calibration of your<br />

rain gauge.<br />

<strong>6700</strong> SAMPLERS<br />

The QUICK VIEW/CHANGE programming style includes a series<br />

of “summary” screens. Summary screens include one or more program<br />

settings. You can quickly step through the program by selecting<br />

the ↓ located in the lower right-hand corner of the display. You<br />

canevenbackstepthroughtheprogrambyselectingthe↑ located in<br />

the upper left hand corner of the display. Program changes are made<br />

by selecting the program setting you wish to change. Then make the<br />

changes as you would in the NORMAL programming mode. The<br />

extended programming mode defaults to QUICK VIEW/CHANGE.<br />

93<br />

HARDWARE SETUP<br />

The hardware setup screens, available with extended programming,<br />

make a number of adjustments to the <strong>Sampler</strong>’s hardwarebased<br />

features:<br />

Liquid Detector Disable Rain Gauge Setup<br />

YSI 600 Sonde Setup Master/slave Sampling<br />

Full Bottle Detection Display Backlighting<br />

Event Marks Purge Counts<br />

I/O Pin Programming Analog Outputs<br />

Serial Output Dialout Alarms<br />

LIQUID DETECTOR DISABLE<br />

Under some conditions (pressurized lines, sampling foam), a more<br />

repeatable sample volume may be delivered by disabling the liquid<br />

detector. Selecting NO will turn the Liquid Detector off.<br />

If the Liquid Detector is off:<br />

you must enter the SUCTION HEAD,<br />

no RINSES or RETRIES will be available,<br />

you will not be able to CALIBRATE SAMPLE volumes,<br />

the RESULTS report will read a COUNT TO LIQUID of 0,<br />

a complete purge will occur between samples when using multiple<br />

bottles-per-sample distribution.<br />

RAIN GAUGE SETUP<br />

With standard programming, the sampler automatically records<br />

the rain gauge readings. However, with extended programming,<br />

recording the rain gauge readings and the rain enable option can<br />

be turned on or off. Enter the hardware setup and follow screens<br />

43 and 44 in the margin.<br />

At screen 43, select YES to record the rain gauge readings and<br />

add RAIN as a sampler enable option. Continue with screen 44<br />

and the sampler will then create a rain memory partition and<br />

update the available sampler enable conditions.<br />

Select NO to disable the readings and the sampler enable option.<br />

When you select NO, the sampler removes the rain memory partition,<br />

leaving the memory space available for other reading types.