Other Functions Menu<br />

RUN “PROGRAM NAME”<br />

PROGRAM<br />

VIEW REPORT<br />

OTHER FUNCTIONS<br />

Select OTHER FUNCTIONS.<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

NotethattheHARDWARE<br />

option appears only in<br />

extended programming.<br />

Operating the Pump<br />

GRAB SAMPLE<br />

CALIBRATE VOLUME<br />

OPERATE PUMP<br />

MOVE DISTRIBUTOR<br />

SELECT DIRECTION<br />

PUMP FORWARD<br />

PUMP REVERSE<br />

PURGING. . .<br />

PRESS THE RED ‘STOP’<br />

KEY WHEN DONE<br />

PURGED _____COUNTS<br />

Moving the Distributor<br />

Arm<br />

GRAB SAMPLE<br />

CALIBRATE VOLUME<br />

OPERATE PUMP<br />

MOVE DISTRIBUTOR<br />

Select MOVE<br />

DISTRIBUTOR<br />

GO TO BOTTLE ____<br />

(1-MAX)<br />

Enter the bottle number<br />

to which you want to<br />

move the distributor<br />

NOW AT BOTTLE ____<br />

<strong>6700</strong> SAMPLERS<br />

OTHER FUNCTIONS<br />

The “Other Functions” menu screen includes options for<br />

Maintenance (See Chapter 4.)<br />

<strong>Manual</strong> Functions.<br />

Programming Style.<br />

Hardware (See Hardware Setup.) The Hardware menu option<br />

appears in extended programming only.<br />

MANUAL FUNCTIONS<br />

The manual functions programming screens let you:<br />

Take grab samples.<br />

Calibrate sample volumes.<br />

Operate the pump manually.<br />

Move the distributor.<br />

Grab samples and calibrate volumes are discussed in Chapter 3.<br />

OPERATING THE PUMP<br />

You can manually operate the sampler’s pump in the forward or<br />

reverse direction. You can do this to test the pump, draw a large<br />

sample, or purge the suction line.<br />

When you run the pump manually, the sampler reports the number<br />

of pump counts during the pump’s operation. Use this feature<br />

to obtain an estimate of the pump counts required to purge a<br />

nonstandard suction line.<br />

MOVING THE DISTRIBUTOR ARM<br />

The sampler lets you reposition the distributor arm. Use this feature<br />

to verify a bottle location when installing a bottle kit. You<br />

cannot move the distributor while the sampler runs a program.<br />

Always use the steps shown in the margin.<br />

DO NOT ROTATE THE DISTRIBUTOR ARM MANUALLY; THIS<br />

WILL SEVERELY DAMAGE THE DISTRIBUTOR DRIVE GEARS.<br />

PROGRAMMING STYLE<br />

The sampler has two different styles of programming screens;<br />

NORMAL and QUICK VIEW/ CHANGE.<br />

The NORMAL programming style steps you through each programming<br />

step. You simply answer each question as it is asked.<br />

The standard programming mode defaults to NORMAL.<br />

92<br />