3. Secure the slide clamp as close to the extension set Luer lock connector<br />

as possible. This should assure a minimum water loss from<br />

the tubing when the syringe is removed.<br />

4. Weigh the entire Medication Cassette Reservoir/extension set and<br />

record the weight. This is the predelivery weight. (This weight<br />

includes the empty Medication Cassette Reservoir, extension set,<br />

and weight of the water.)<br />

5. Attach the cassette to the pump. Program the Reservoir Volume to<br />

20 ml. Now press ¤. This value is the intended delivery volume.<br />

(1 ml of water at 20°C weighs 1 gram.) Remove the slide clamp.<br />

6. With the pump in Lock Level 0, program a Continuous Rate of 125<br />

ml/hr. Start the pump and wait for the reservoir volume empty<br />

alarm. The pump will deliver 20 ml.<br />

7. Again, secure the slide clamp as close as possible to end of the<br />

extension set Luer lock connector. Remove the cassette from the<br />

pump and weigh the entire Medication Cassette Reservoir/extension<br />

set assembly. This is the postdelivery weight.<br />

8. Calculate the difference in weight between the predelivery weight<br />

and the postdelivery weight. This is the weight of the amount<br />

delivered.<br />

9. Find the difference between the volume of the amount delivered<br />

and the intended delivery volume. This is the inaccuracy volume.<br />

10. Divide the inaccuracy volume by the intended delivery volume and<br />

multiply by 100. This is the accuracy error percentage.<br />

11. If the accuracy error percentage is greater than ± 6%, repeat the<br />

test with a new Medication Cassette Reservoir. If the pump fails<br />

a second time, call <strong>Smiths</strong> <strong>Medical</strong> MD, Inc. or <strong>Smiths</strong> <strong>Medical</strong><br />