Reference<br />

Section 5: Reference<br />

0<br />

6. Slowly increase the backpressure, noting when the high pressure<br />

alarm is activated.<br />

NOTE: The pressure may be increased rapidly to 8 psi [0.55 bar],<br />

after which the pressure should be increased at 3 psi/min [0.21 bar/<br />

min] or less until the alarm sounds.<br />

7. The high pressure alarm should sound between 12 and 40 psi<br />

(26 ± 14 psi) [between 0.82 and 2.76 bar (1.79 ± 0.97 bar)].<br />

CAUTION: At the completion of the test, the pressure must be<br />

reduced to zero before detaching the cassette from the pump;<br />

otherwise, the cassette may rupture. Safety glasses should be<br />

worn while conducting or observing this test.