Upstream Occlusion Sensor Test<br />

Section 5: Reference<br />

• Verify the Upstream Occlusion Sensor is turned On (see Section 4,<br />

Biomed Functions).<br />

• Obtain a CADD Administration Set with bag spike. Also obtain a<br />

clamp (slide clamp or hemostat).<br />

• Insert the CADD Administration Set spike into an appropriate,<br />

standard IV bag filled with water. Attach the cassette to the pump.<br />

Prime the entire fluid path.<br />

• Program the pump to deliver a continuous maximum rate. Press and<br />

hold ⁄ to start the pump.<br />

• Clamp the tubing halfway between the IV bag and the pump. The<br />

pump should alarm within 3 activations after clamping the tubing.<br />