Upstream Occlusion Sensor Test<br />
Section 5: Reference<br />
• Verify the Upstream Occlusion Sensor is turned On (see Section 4,<br />
Biomed Functions).<br />
• Obtain a CADD Administration Set with bag spike. Also obtain a<br />
clamp (slide clamp or hemostat).<br />
• Insert the CADD Administration Set spike into an appropriate,<br />
standard IV bag filled with water. Attach the cassette to the pump.<br />
Prime the entire fluid path.<br />
• Program the pump to deliver a continuous maximum rate. Press and<br />
hold ⁄ to start the pump.<br />
• Clamp the tubing halfway between the IV bag and the pump. The<br />
pump should alarm within 3 activations after clamping the tubing.<br />