Reference<br />

Section 5: Reference<br />

Cassette Sensor Check<br />

• Unlock the cassette by inserting a coin into the lock and turning<br />

clockwise.<br />

• The pump will sound a continuous two-tone alarm and the display<br />

should show “No Disposable, Clamp Tubing.”<br />

• Press ⁄ or „ to silence the alarm. Press and hold Å to turn<br />

the pump off.<br />

The following 3 checks (LCD, Motor and Gear Train and Reservoir Volume<br />

Empty Alarm Check) should be performed in the sequence shown.<br />

LCD Check<br />

• With the pump turned off, press Å. The second screen that the<br />

pump displays will consist of 32 blocks of characters. Examine the<br />

LCD to verify that there are no missing pixels (dark dots) in the character<br />

blocks.<br />

Motor and Gear Train Check<br />

• Program the Reservoir Volume to 2.0 ml.<br />

• Attach a 50 or 100 ml Medication Cassette Reservoir or CADD Administration<br />

Set to the pump. Lock the cassette.<br />

• Press and hold ‹ until 3 series of dashes appear. Release ‹.<br />

Press and hold ‹. While priming the tubing, listen to the motor<br />

for excessive noise or grinding sounds. Count the number of pump<br />

activations. The pump should prime 10 double activations and then<br />

stop. Press „ to return to the main menu.<br />

Reservoir Volume Empty Alarm Check<br />

• Program the Reservoir Volume to 1.0 ml. Press „ until Reservoir<br />

Volume is displayed on the LCD. Press ´ or Î until 1.0 ml is<br />

displayed. Then press ¤.