Section 5: Reference<br />
the motor drive circuit without running the motor. The microprocessor<br />
performs this test function every several minutes to assure its continued<br />
functionality. An input from the watchdog circuit prevents motor operation<br />
if the watchdog timer expires. The software verifies this function<br />
during the watchdog test described above.<br />
Voltage Detector Circuit<br />
Low voltage detection is performed by part of the Watchdog Circuit and<br />
by the microprocessor via software. Three low voltage levels are detected.<br />
The first 2 levels are detected by software and the third by hardware.<br />
The first level to be reached is the Low Battery Warning threshold which<br />
occurs when the battery voltage decays to a nominal value of 2.4 volts<br />
when the motor is off or 1.8 volts when the motor is active. An Analog<br />
to Digital Converter (ADC) built into the microprocessor allows the<br />
microprocessor, via software, to monitor the battery voltage. At the Low<br />
Battery Warning threshold, the microprocessor enables a periodic series<br />
of beeps and displays a low battery warning message on the LCD. As the<br />
voltage operating the motor reaches a nominal value of 4.75 volts, the<br />
software disables delivery, places a battery depleted message on the LCD,<br />
and enables a constant two tone audible alarm. When the battery voltage<br />
decays to a nominal value of 1.0 volts, a hardware reset circuit is triggered<br />
which places the microprocessor in reset. This prevents ambiguous<br />
microprocessor operation when the battery voltage continues to decay.<br />
The hardware reset continues until the battery is completely discharged<br />
or until it is removed. Once the pump controller goes into low battery<br />
shutdown, only replacing the depleted batteries with new ones will clear<br />
the condition.<br />