Reference<br />

Section 5: Reference<br />

Accuracy Test Results<br />

The following graphs are designed to show flow accuracy of the infusion<br />

system plotted against given time periods.<br />

Flow rate: Intermediate<br />

Time Interval: 0.5 min<br />

Total Time: 120 min<br />

Programmed Rate: 24.0 ml/hr<br />

Flow<br />

(ml/hr)<br />

48<br />

36<br />

Flow<br />

(ml/hr)<br />

24<br />

12<br />

0<br />

0 10 20 30 40 50 60 70 80 90 100 110 120<br />

T (min)<br />

Trumpet curve: Intermediate rate<br />

Programmed Rate: 24.0 ml/hr<br />

Average Flow Rate: 23.9227 ml/hr<br />

Mean Flow Error: -0.32%<br />

% Error<br />

of Flow<br />

T (min)<br />

T (min)