• Do not expose the pump to humidity levels below 20% or above 90%<br />

relative humidity.<br />

• Do not store the pump for prolonged periods of time with the batteries<br />

installed.<br />

• Frozen medication must be thawed at room temperature only. Do not<br />

heat the Medication Cassette Reservoir in a microwave oven as this<br />

may damage the medication, the Medication Cassette Reservoir, or<br />

cause leakage.<br />

• Do not immerse the pump in cleaning fluid or water or allow solution<br />

to soak into the pump, accumulate on the keypad, or enter the battery<br />

compartment.<br />

• Do not clean the pump with acetone, other plastic solvents, or abrasive<br />

cleaners, as damage to the pump may occur.<br />

• Do not expose the pump to therapeutic levels of ionizing radiation<br />

as permanent damage to the pump’s electronic circuitry may occur.<br />

The best procedure to follow is to remove the pump from the patient<br />

during therapeutic radiation sessions. If the pump must remain in the<br />

vicinity during a therapy session, it should be shielded, and its ability<br />

to function properly should be confirmed following treatment.<br />

• Do not expose the pump directly to ultrasound, as permanent damage<br />

to the pump’s electronic circuitry may occur.<br />

• Do not use the pump in the vicinity of magnetic resonance imaging<br />

(MRI) equipment as magnetic fields may adversely affect the operation<br />

of the pump. Remove the pump from the patient during MRI procedures<br />

and keep it at a safe distance from magnetic energy.<br />

• Do not use the pump near ECG equipment as the pump may interfere<br />

with the operation of the equipment. Monitor ECG equipment carefully<br />

when using this pump.<br />

• Do not sterilize the pump.<br />

• Use only <strong>Smiths</strong> <strong>Medical</strong> MD, Inc. accessories as using other brands<br />

may adversely affect the operation of the pump.<br />