Section 5: Reference<br />

System Delivery<br />

Accuracy .............................± 6% (nominal). At low infusion rates, this accuracy<br />

may not be achieved for short periods. During<br />

the total infusion time, the accuracy averages<br />

out (see accuracy curves, pages 62 and 63)<br />

WARNING: Ensure that the ± 6% System Delivery Accuracy specification<br />

is taken into account when programming the pump and/or filling the<br />

Medication Cassette Reservoir. Failure to do so may result in medication<br />

in the reservoir becoming depleted sooner than expected. If the pump is<br />

being used to deliver critical or life sustaining medication, the interruption<br />

in the delivery of medication could result in patient injury or death.<br />

System Definition ..............System is defined as a CADD-<strong>Legacy®</strong> pump<br />

with an attached Medication Cassette Reservoir<br />

and CADD Extension Set with integral antisiphon<br />

valve, or an attached CADD Administration<br />

Set with integral or add-on anti-siphon<br />

valve. OR a CADD-<strong>Legacy®</strong> pump with an<br />

attached Medication Cassette Reservoir with<br />

flow stop feature and CADD Extension Set,<br />

or a CADD Administration Set with flow stop<br />

feature (catalog numbers start with 21-73xx).<br />

High Pressure Alarm ........26 ± 14 psi [1.79 ± 0.97 bar].<br />

Air Detector Alarm ...........Single bubble<br />

Low sensitivity = greater than 0.250 ml<br />

High Sensitivity = greater than 0.100 ml<br />

Multi-bubble = 1.0 ml nominal<br />

Maximum Volume<br />

Infused under Single<br />

Fault Condition .................CADD Administration Set: 0.2 ml.<br />

Delivery Rate during<br />

priming ...............................Approx. 180 ml/hr<br />

Alarm Disabled during<br />

priming ...............................Air Detector<br />

Reference