Section 5: Reference<br />

Exposure to Radiation, Ultrasound, Magnetic<br />

Resonance Imaging (MRI), or Use near ECG Equipment<br />

CAUTION:<br />

• Do not expose the pump to therapeutic levels of ionizing radiation<br />

as permanent damage to the pump’s electronic circuitry may<br />

occur. The best procedure to follow is to remove the pump from the<br />

patient during therapeutic radiation sessions. If the pump must remain<br />

in the vicinity during a therapy session, it should be shielded,<br />

and its ability to function properly should be confirmed following<br />

treatment.<br />

• Do not expose the pump directly to ultrasound, as permanent damage<br />

to the pump’s electronic circuitry may occur.<br />

• Do not use the pump in the vicinity of magnetic resonance imaging<br />

(MRI) equipment as magnetic fields may adversely affect the operation<br />

of the pump. Remove the pump from the patient during MRI<br />

procedures and keep it at a safe distance from magnetic energy.<br />

• Do not use the pump near ECG equipment as the pump may interfere<br />

with the operation of the equipment. Monitor ECG equipment<br />

carefully when using this pump.<br />