Reference<br />
Section 5: Reference<br />
Messages and Alarms Description/Corrective Action<br />
High Pressure<br />
Two-Tone alarm<br />
Key pressed,<br />
Please release<br />
Two-Tone alarm<br />
LowBat<br />
Three Two-Tone beeps<br />
every 5 minuTes<br />
Motor Locked,<br />
remove all power<br />
Two-Tone alarm<br />
[No message]<br />
Two-Tone alarm<br />
0<br />
The pump has detected high pressure,<br />
which may be resulting from a<br />
downstream blockage, kink in the fluid<br />
path, or a closed tubing clamp. Remove<br />
the obstruction to resume operation.<br />
Or, press „ or ⁄ to stop the pump<br />
and silence the alarm for 2 minutes, then<br />
remove the obstruction and restart the<br />
pump.<br />
If a key is being pressed, stop pressing<br />
it. If the alarm persists, close the tubing<br />
clamp and remove the pump from use.<br />
Contact Customer Service to return the<br />
pump for service.<br />
The batteries are low, but the pump is<br />
still operable.<br />
• Change the batteries soon.<br />
Batteries are depleted and the pump was<br />
powered up with the AC Adapter. Install<br />
new AA batteries, reconnect the AC<br />
adapter, and restart the pump.<br />
With no AC adapter attached, the batteries<br />
have been removed while the pump is<br />
running. The pump is now stopped and<br />
unpowered. Install batteries to silence the<br />
alarm.<br />
OR<br />
Batteries were removed within<br />
approximately 15 seconds after stopping the<br />
pump. Install new batteries to silence the<br />
alarm, if desired. Otherwise, the alarm will<br />
stop within a short period of time.