5.0 Reference<br />

Messages and Alarms, Alphabetical List<br />

Section 5: Reference<br />

Messages and Alarms Description/Corrective Action<br />

Air In Line Detected<br />

Two-Tone alarm<br />

Battery Depleted<br />

Two-Tone alarm<br />

Battery Removed<br />

Pump won’t run<br />

Two-Tone alarm<br />

Error<br />

Two-Tone alarm<br />

The Air Detector has detected air in the<br />

fluid path; the fluid path may contain air<br />

bubbles, or the tubing may not be fully<br />

threaded through the Air Detector. Press<br />

⁄ or „ to silence the alarm, then:<br />

• Make sure the tubing is threaded<br />

properly.<br />

• If the fluid path contains air bubbles,<br />

close the clamps and disconnect the fluid<br />

path from the patient. Then follow the<br />

instructions for removing air by priming<br />

the pump, described in Section 2. Restart<br />

the pump.<br />

The battery power is too low to operate the<br />

pump. The pump is now stopped.<br />

• Change the batteries immediately.<br />

• Press and hold ⁄ to restart the pump.<br />

With the AC adapter attached, the AA<br />

batteries have been removed while the pump<br />

is running, or you have tried to start the<br />

pump with depleted batteries. The pump is<br />

now stopped. Press ⁄ or „ to silence<br />

the alarm. Reinstall batteries or install new<br />

batteries. Press and hold ⁄ to restart the<br />

pump.<br />

An error has occurred. Remove the pump<br />

from service and contact Customer Service<br />

to return the pump for service.<br />