Biomed<br />
Functions<br />
Section 4: Biomed Functions<br />
Changing the Delivery Method<br />
Delivery can be changed between Continuous Delivery and Intermittent<br />
Delivery. This allows for either continuous medication delivery or<br />
infusion of a prescribed volume of a medication over a specified time<br />
period. When the delivery method is changed, programming will revert<br />
to default settings.<br />
1. With the pump stopped and in LL0, access Biomed Functions.<br />
(Refer to the beginning of the Biomed Functions section for instructions<br />
on how to access Biomed Functions.)<br />
2. Press „ until “Delivery Mode” appears.<br />
3. Press ´ or Î to select “Continuous” or “Intermittent.”<br />
4. Press ¤ to enter the change.