Biomed<br />

Functions<br />

Section 4: Biomed Functions<br />

Changing the Delivery Method<br />

Delivery can be changed between Continuous Delivery and Intermittent<br />

Delivery. This allows for either continuous medication delivery or<br />

infusion of a prescribed volume of a medication over a specified time<br />

period. When the delivery method is changed, programming will revert<br />

to default settings.<br />

1. With the pump stopped and in LL0, access Biomed Functions.<br />

(Refer to the beginning of the Biomed Functions section for instructions<br />

on how to access Biomed Functions.)<br />

2. Press „ until “Delivery Mode” appears.<br />

3. Press ´ or Î to select “Continuous” or “Intermittent.”<br />

4. Press ¤ to enter the change.