Operating<br />

the Pump<br />

Section 3: Operating the Pump<br />

Turning the Pump On/Off<br />

When the pump is stopped, you may put the pump into a low power state<br />

by turning it off. The pump may be turned off when it is disconnected<br />

from the patient and it is going to be stored for short periods of time.<br />

CAUTION: Do not store the pump for prolonged periods of time with<br />

the batteries installed. Battery leakage could damage the pump.<br />

To turn the pump off<br />

• Press and hold Å.<br />

Three sets of dots will appear one-by-one on the pump’s display,<br />

each accompanied by a single beep.<br />

To turn the pump on<br />

• Press and hold Å. The pump will power up and automatically<br />

review all screens.<br />

Resetting the Reservoir Volume<br />

To reset the Reservoir Volume to the value programmed in LL0, the<br />

pump may be in any lock level.<br />

1. Stop the pump.<br />

2. Press „ to display the Reservoir Volume screen.<br />

3. Press ¤ to reset the volume to the programmed value.