Section 2: Pump Setup and Programming<br />
IDM: Stopping the Pump During the Dose<br />
Stopping the pump while a dose is in progress will shift all subsequent<br />
doses by the amount of time the pump is stopped.<br />
If you wish to make up for the lost time in the cycle, it is recommended<br />
that you wait until the current dose is completed. If the pump is stopped<br />
while a dose is infusing and you reprogram the “Dose Starts in” screen,<br />
then the pump will not deliver the remainder of that dose. (To determine<br />
the amount of time remaining in the current dose, use the Dose Remaining<br />
screen in the programming loop.) Then stop the pump, change to LL0<br />
or LL1, and reprogram the delay on the “Dose Starts in” screen.<br />
Resuming the Dose<br />
To resume delivery of a dose that has been stopped in progress, simply<br />
restart the pump.<br />
Resetting the Cycle<br />
Pump is<br />
stopped for<br />
10 minutes<br />
All<br />
subsequent<br />
doses shift by<br />
10 minutes<br />
Dose Dose<br />
If you wish to reset the cycle, stop the pump during KVO, change to LL0<br />
or LL1, and reprogram the start of the Next Dose.<br />
If a dose has been stopped in progress and you wish to cancel the current<br />
dose and reset the cycle, make sure the pump is in Lock Level 0 or Lock<br />
Level 1. The Dose Remaining screen with the remaining hours and minutes<br />
of the dose will be displayed, followed by the “Dose Starts in” screen.<br />
Reprogram the “Dose Starts in” screen, which will cancel the remainder of<br />
the dose, and the next cycle will start after the delay you selected.<br />
Pump Setup &<br />
Programming