Pump Setup &<br />

Programming<br />

Section 2: Pump Setup and Programming<br />

CDM: Programming with Upper Limits, Adjusting Rate<br />

in Lock Level 1<br />

This feature may be used in the continuous delivery mode only. If a<br />

prescription allows for the Continuous Rate to be adjusted during the<br />

course of therapy, you may wish to operate the pump in LL1. Then, when<br />

necessary, you can adjust the Continuous Rate up to the maximum value<br />

that was programmed in LL0.<br />

Programming the pump to use this feature<br />

0<br />

1. During initial programming in LL0, enter the upper limit value<br />

for the Continuous Rate. (This will be the maximum value when<br />

the pump is in LL1.)<br />

2. After you are finished programming, change the lock level to LL1.<br />

3. Decrease the Continuous Rate to its starting value, then press ¤.<br />

Adjusting the rate while the pump is in use<br />

If it becomes necessary to increase the Continuous Rate during the<br />

course of therapy, stop the pump but remain in LL1.<br />

1. Press „ until the Continuous Rate screen appears.<br />

2. Press ´ or Î to select the desired value, then press ¤.<br />

3. Restart the pump if appropriate.