Pump Setup &<br />
Programming<br />
Section 2: Pump Setup and Programming<br />
CDM: Programming with Upper Limits, Adjusting Rate<br />
in Lock Level 1<br />
This feature may be used in the continuous delivery mode only. If a<br />
prescription allows for the Continuous Rate to be adjusted during the<br />
course of therapy, you may wish to operate the pump in LL1. Then, when<br />
necessary, you can adjust the Continuous Rate up to the maximum value<br />
that was programmed in LL0.<br />
Programming the pump to use this feature<br />
0<br />
1. During initial programming in LL0, enter the upper limit value<br />
for the Continuous Rate. (This will be the maximum value when<br />
the pump is in LL1.)<br />
2. After you are finished programming, change the lock level to LL1.<br />
3. Decrease the Continuous Rate to its starting value, then press ¤.<br />
Adjusting the rate while the pump is in use<br />
If it becomes necessary to increase the Continuous Rate during the<br />
course of therapy, stop the pump but remain in LL1.<br />
1. Press „ until the Continuous Rate screen appears.<br />
2. Press ´ or Î to select the desired value, then press ¤.<br />
3. Restart the pump if appropriate.