Setting the Lock Level for the Patient<br />

Section 2: Pump Setup and Programming<br />

The Lock Level must be changed to LL1 or LL2 to prevent the patient from<br />

having complete access to all programming and operating functions.<br />

NOTE: You may change the lock level at any time by stopping the pump<br />

and following the procedure below.<br />

To change the lock level<br />

1. Press Œ.<br />

2. The current lock level will appear.<br />

3. Press ´ or Î until the desired lock level (LL1 or LL2) appears.<br />

4. Press Œ again or ¤. “Code 0” will appear.<br />

5. Press ´ or Î until ** Text Omitted **<br />

6. Press Œ or ¤ to set the new lock level.<br />

WARNING: Do not disclose to the patient the pump’s security<br />

codes or any other information that would allow the patient<br />

complete access to all programming and operating functions.<br />

Improper programming could result in death or serious injury<br />

to the patient.<br />