Pump Setup &<br />
Programming<br />
Section 2: Pump Setup and Programming<br />
3. To seat the tubing into the groove, gently pull the tubing, until it is<br />
under the retention nubs and flat in the groove.<br />
Retent on<br />
nubs<br />
4. Connect the tubing to the patient’s infusion set or indwelling<br />
catheter.<br />
WARNING: Ensure that the entire fluid path is free of all air<br />
bubbles before connecting to the patient to prevent air embolism.<br />
Air embolism could result in death or serious injury to the<br />
patient.<br />
NOTE: If the fluid path contains an air eliminating filter, it is acceptable<br />
for air bubbles to be present on the vent side of the filter.