Section 2: Pump Setup and Programming<br />
6. If the Air Detector is in use, go to the next section. If not, connect<br />
the tubing to the patient’s infusion set or indwelling catheter and<br />
go to Setting the Lock Level for the Patient.<br />
WARNING: Ensure that the entire fluid path is free of all air bubbles<br />
before connecting to the patient to prevent air embolism. Air<br />
embolism could result in death or serious injury to the patient.<br />
NOTE: If the fluid path contains an air eliminating filter, it is<br />
acceptable for air bubbles to be present on the vent side of the filter.