Pump Setup &<br />

Programming<br />

Section 2: Pump Setup and Programming<br />

6. If the Air Detector is in use, go to the next section. If not, connect<br />

the tubing to the patient’s infusion set or indwelling catheter and<br />

go to Setting the Lock Level for the Patient.<br />

WARNING: Ensure that the entire fluid path is free of all air bubbles<br />

before connecting to the patient to prevent air embolism. Air<br />

embolism could result in death or serious injury to the patient.<br />

NOTE: If the fluid path contains an air eliminating filter, it is<br />

acceptable for air bubbles to be present on the vent side of the filter.