14.03.2013 Views

Legacy® PLUS - Smiths Medical

Legacy® PLUS - Smiths Medical

Legacy® PLUS - Smiths Medical

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 2: Pump Setup and Programming<br />

Priming the Tubing (Using the Pump) and Connecting<br />

to the Patient<br />

The pump must be stopped and in LL0 or LL1 in order to prime the fluid<br />

path. If the pump is in LL2, you cannot prime the fluid path.<br />

NOTE: If you are not changing the fluid container but wish to prime the<br />

fluid path, you may follow the same procedure.<br />

WARNING: Do not prime the fluid path with the tubing connected<br />

to a patient as this could result in overdelivery of medication or air<br />

embolism, which could result in death or serious injury to the patient.<br />

1. Make sure the tubing is disconnected from the patient and the tubing<br />

clamp is open.<br />

2. Press and hold ‹. You will hear a single beep, and the word<br />

“Prime” will appear on the display.<br />

3. After “Prime” and 3 sets of dashes appear, and you hear 3 beeps,<br />

release ‹.<br />

4. Press and hold ‹ again to fill the fluid path and to eliminate<br />

air bubbles. The screen displays “Priming . . .” and you will hear a<br />

short beep each time the pump goes through a delivery cycle.<br />

NOTE:<br />

• The air detector alarm is automatically disabled when priming.<br />

• Fluid delivered during priming is subtracted from the Reservoir<br />

Volume, but is not added to the Given screen since this fluid is not<br />

delivered to the patient.<br />

5. If the tubing is not yet fully primed, press and hold ‹ again. If<br />

the tubing is primed, press „ to return to the main screen.<br />

NOTE: Each time you press and hold ‹, you pump a maximum<br />

of 1.0 ml of fluid into the tubing. The pumping action will stop automatically<br />

when 1.0 ml has been delivered. If all of the air has not been<br />

removed from the fluid path, repeat the above priming procedure.<br />

Pump Setup &<br />

Programming

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!