Section 2: Pump Setup and Programming<br />
Priming the Tubing (Using the Pump) and Connecting<br />
to the Patient<br />
The pump must be stopped and in LL0 or LL1 in order to prime the fluid<br />
path. If the pump is in LL2, you cannot prime the fluid path.<br />
NOTE: If you are not changing the fluid container but wish to prime the<br />
fluid path, you may follow the same procedure.<br />
WARNING: Do not prime the fluid path with the tubing connected<br />
to a patient as this could result in overdelivery of medication or air<br />
embolism, which could result in death or serious injury to the patient.<br />
1. Make sure the tubing is disconnected from the patient and the tubing<br />
clamp is open.<br />
2. Press and hold ‹. You will hear a single beep, and the word<br />
“Prime” will appear on the display.<br />
3. After “Prime” and 3 sets of dashes appear, and you hear 3 beeps,<br />
release ‹.<br />
4. Press and hold ‹ again to fill the fluid path and to eliminate<br />
air bubbles. The screen displays “Priming . . .” and you will hear a<br />
short beep each time the pump goes through a delivery cycle.<br />
NOTE:<br />
• The air detector alarm is automatically disabled when priming.<br />
• Fluid delivered during priming is subtracted from the Reservoir<br />
Volume, but is not added to the Given screen since this fluid is not<br />
delivered to the patient.<br />
5. If the tubing is not yet fully primed, press and hold ‹ again. If<br />
the tubing is primed, press „ to return to the main screen.<br />
NOTE: Each time you press and hold ‹, you pump a maximum<br />
of 1.0 ml of fluid into the tubing. The pumping action will stop automatically<br />
when 1.0 ml has been delivered. If all of the air has not been<br />
removed from the fluid path, repeat the above priming procedure.<br />
Pump Setup &<br />
Programming