Section 2: Pump Setup and Programming<br />
Removing a Cassette<br />
WARNING: Close the fluid path tubing with the clamp before removing<br />
the cassette from the pump to prevent unregulated gravity infusion, which<br />
could result in death or serious injury to the patient.<br />
1. Stop the pump.<br />
2. Close the tubing clamp.<br />
3. Insert a coin into the lock and turn it clockwise. The lock will pop<br />
out when you unlock the cassette.<br />
4. A continuous alarm will sound and the pump will display<br />
“No Disposable, Clamp Tubing.” The alarm may be silenced by<br />
pressing ⁄ or „.<br />
5. Remove the cassette hooks from the pump hinge pins.