Pump Setup &<br />

Programming<br />

Section 2: Pump Setup and Programming<br />

Removing a Cassette<br />

WARNING: Close the fluid path tubing with the clamp before removing<br />

the cassette from the pump to prevent unregulated gravity infusion, which<br />

could result in death or serious injury to the patient.<br />

1. Stop the pump.<br />

2. Close the tubing clamp.<br />

3. Insert a coin into the lock and turn it clockwise. The lock will pop<br />

out when you unlock the cassette.<br />

4. A continuous alarm will sound and the pump will display<br />

“No Disposable, Clamp Tubing.” The alarm may be silenced by<br />

pressing ⁄ or „.<br />

5. Remove the cassette hooks from the pump hinge pins.