11. Verify the Upstream Sensor status.<br />

Section 2: Pump Setup and Programming<br />

• Make sure the desired setting is displayed. This screen will show<br />

whether the Upstream Occlusion Sensor is turned on or off.<br />

WARNING: When the Upstream Occlusion Sensor is turned off,<br />

the pump will not detect occlusions upstream (between pump and<br />

fluid container). Periodically inspect the fluid container for decreasing<br />

volume, inspect the fluid path for kinks, a closed clamp,<br />

or other upstream occlusions. Upstream occlusions could result in<br />

under- or non-delivery of medications. If undetected, these occlusions<br />

could result in death or serious injury to the patient.<br />

• If you need to change the Upstream Occlusion Sensor setting, see<br />

Section 4, Biomed Functions.<br />

• Press „.<br />

12. Review the program.<br />

Press „ repeatedly to review the programming screens. If you<br />

need to reprogram a setting, press „ until the appropriate<br />

screen appears and change the setting as described in this section.<br />

Pump Setup &<br />

Programming