Programming Intermittent Delivery<br />

Section 2: Pump Setup and Programming<br />

WARNING: This pump delivers medication at a continuous rate or<br />

intermittent rate in milliliters per hour (ml/hr). Programming the<br />

pump at a delivery rate other than what is prescribed will cause over<br />

or under delivery of medication, which could result in patient injury<br />

or death.<br />

Be sure the pump is in the intermittent delivery mode versus the continuous<br />

delivery mode. To program the pump, enter the prescribed values.<br />

1. Begin at the main screen.<br />

• Make sure the pump is in LL0.<br />

• Make sure STOPPED appears on the main screen.<br />

• Press „ to begin.<br />

2. Enter the Reservoir Volume.<br />

• Press ´ or Î to select the volume of a filled fluid container.<br />

(If you do not wish to use the Reservoir Volume feature, scroll<br />

down to “Not In Use” located before 1.)<br />

• Press ¤.<br />

• Press „.<br />

3. Enter the Dose Volume.<br />

• Press ´ or Î to select the desired volume.<br />

• Press ¤.<br />

• Press „.<br />

NOTE: The maximum rate of the pump is 125 ml/hr. The pump will<br />

only allow a dose duration that will not cause a rate greater than<br />

125 ml/hr for the programmed volume.<br />

4. Enter the Dose Duration.<br />

• Press ´ or Î to select the desired duration.<br />

• Press ¤.<br />

• Press „.<br />

Pump Setup &<br />

Programming