Pump Setup &<br />

Programming<br />

Section 2: Pump Setup and Programming<br />

Air Detector<br />

This screen indicates whether the Air Detector is on high sensitivity,<br />

low sensitivity or turned off. The Air Detector status cannot be changed<br />

without entering the Biomed Functions Code (see Section 4, Biomed<br />

Functions, to change the setting).<br />

Upstream Sensor Status<br />

This screen indicates whether the Upstream Occlusion Sensor is turned<br />

on or turned off. The Upstream Sensor status cannot be changed without<br />

entering the Biomed Functions Code (see Section 4, Biomed Functions,<br />

to change the setting).<br />

Before programming in Intermittent Delivery Mode:<br />

• Stop the pump and change the Lock Level to LL0 as described in Section 1.<br />

• Select the Intermittent Delivery Mode (IDM) as described in Changing<br />

the Delivery Method in Section 4. If the pump is already in the<br />

IDM, you may select it again to clear all programming screens to their<br />

default settings.