Section 2: Pump Setup and Programming<br />

Continuous Rate<br />

Enter the continuous rate of medication delivery (in ml/hr). The maximum<br />

rate is 125 ml/hr.<br />

Given<br />

This screen shows the total amount of medication delivered since the last<br />

time the value was cleared. The amount shown is rounded to the nearest<br />

0.05 ml. If this value reaches 99999.95, it automatically returns to 0 and<br />

continues counting. The Given amount does not include medication used<br />

when priming the tubing.<br />

Air Detector Status<br />

This screen indicates whether the Air Detector is on high sensitivity, low<br />

sensitivity or turned off. The Air Detector status cannot be changed without<br />

entering the Biomed Functions Code (see Section 4, Biomed Functions,<br />

to change the setting).<br />

Upstream Sensor Status<br />

This screen indicates whether the Upstream Occlusion Sensor is turned<br />

on or turned off. The Upstream Sensor status cannot be changed without<br />

entering the Biomed Functions Code (see Section 4, Biomed Functions, to<br />

change the setting).<br />

Before programming in Continuous Delivery Mode:<br />

• Stop the pump and change the Lock Level to LL0 as described in Section 1.<br />

• Select the Continuous Delivery Mode (CDM) as described in Changing<br />

the Delivery Methods in Section 4. If the pump is already in the<br />

CDM, you may select it again to clear all programming screens to<br />

their default settings.