Section 2: Pump Setup and Programming<br />

Programming Screens for Continuous Delivery<br />

These are the programming screens for the CADD-<strong>Legacy®</strong> <strong>PLUS</strong> pump,<br />

in the continuous delivery mode. Descriptions of the screens follow.<br />

Reservoir Volume<br />

Continuous Rate<br />

(ml/hr)<br />

Given<br />

Air Detector (Off,<br />

On-High, or On-Low)<br />

Upstream Sensor<br />

(Off or On)<br />

Reservoir Volume<br />

Reservoir Volume<br />

100.0 ml<br />

Continuous Rate<br />

5.0 ml/hr<br />

Given<br />

2.50 ml<br />

Air Detector<br />

On-High<br />

Upstream Sensor<br />

On<br />

Enter the volume of fluid contained in a filled fluid container. The Reservoir<br />

Volume value decreases as the pump delivers fluid or as you prime<br />

the tubing. When you change the fluid container, reset the reservoir<br />

volume on this screen. If you do not wish to use the Reservoir Volume<br />

feature, scroll down to “Not In Use” (located before 1 and after 9999 in<br />

the range of values).<br />

The residual volume could be set higher than the capacity of the fluid<br />

container. Be sure to program the reservoir volume to reflect the actual<br />

volume of the medication being used.<br />