Pump Setup &<br />

Programming<br />

Section 2: Pump Setup and Programming<br />

Programming the Pump: General Instructions<br />

The procedure for changing a programmed setting is similar for most<br />

programming screens.<br />

WARNING: Ensure that the ± 6% System Delivery Accuracy specification<br />

is taken into account when programming the pump and/or filling<br />

the Medication Cassette Reservoir. Failure to do so may result in<br />

medication in the reservoir becoming depleted sooner than expected.<br />

If the pump is being used to deliver critical or life sustaining medication,<br />

the interruption in the delivery of medication could result in<br />

patient injury or death.<br />

• Make sure the pump is stopped and in Lock Level 0.<br />

• To begin programming, start at the main screen and press „.<br />

• To change a setting, press ´ or Î until the desired setting appears.<br />

(Press and hold these keys to change values with increasing<br />

speed.)<br />

• Press ¤ within 25 seconds to confirm a change or the screen will<br />

revert to the previous setting.<br />

• If any key other than ¤ is pressed, “Value not saved” will appear.<br />

Press „ to return to the screen being programmed, scroll to the<br />

desired value, and press ¤.<br />

• Press „ to advance to the next screen.<br />

• To leave a setting unchanged, press „ to go to the next screen.