Pump Setup &<br />
Programming<br />
Section 2: Pump Setup and Programming<br />
Programming the Pump: General Instructions<br />
The procedure for changing a programmed setting is similar for most<br />
programming screens.<br />
WARNING: Ensure that the ± 6% System Delivery Accuracy specification<br />
is taken into account when programming the pump and/or filling<br />
the Medication Cassette Reservoir. Failure to do so may result in<br />
medication in the reservoir becoming depleted sooner than expected.<br />
If the pump is being used to deliver critical or life sustaining medication,<br />
the interruption in the delivery of medication could result in<br />
patient injury or death.<br />
• Make sure the pump is stopped and in Lock Level 0.<br />
• To begin programming, start at the main screen and press „.<br />
• To change a setting, press ´ or Î until the desired setting appears.<br />
(Press and hold these keys to change values with increasing<br />
speed.)<br />
• Press ¤ within 25 seconds to confirm a change or the screen will<br />
revert to the previous setting.<br />
• If any key other than ¤ is pressed, “Value not saved” will appear.<br />
Press „ to return to the screen being programmed, scroll to the<br />
desired value, and press ¤.<br />
• Press „ to advance to the next screen.<br />
• To leave a setting unchanged, press „ to go to the next screen.