General<br />

Description<br />

Section 1: General Description<br />

Lock Level Table<br />

This table lists the operations that are accessible in each lock level while<br />

the pump is stopped and running. LL0 permits complete access to all<br />

programming and operating functions. LL1 permits limited control of<br />

pump programming and operations. LL2 permits only minimal control<br />

of pump operations.<br />

*Continuous Delivery Mode = CDM; Intermittent Delivery Mode = IDM<br />

Pump Operations<br />

and Programming<br />

0<br />

Stopped Running<br />

LL0 LL1 LL2<br />

Any Lock<br />

Level<br />

Stop/Start the pump Yes Yes Yes Yes<br />

Reset Reservo r Volume Yes Yes Yes No<br />

Pr me Yes Yes No No<br />

Change the Lock Level<br />

Change Cont nuous Rate<br />

(CDM)*<br />

Clear G ven amount<br />

(CDM & IDM)*<br />

Yes, w/<br />

code<br />

Yes<br />

Yes, w/<br />

code<br />

Up to LL0<br />

value<br />

Yes, w/<br />

code<br />

No<br />

No No<br />

Yes Yes No No<br />

Change Dose Volume (IDM)* Yes No No No<br />

Change Dose Durat on (IDM)* Yes No No No<br />

Change Dose Cycle (IDM)* Yes No No No<br />

Change KVO Rate (IDM)* Yes No No No<br />

Change Dose Starts n Delay<br />

(IDM)*<br />

Biomed Functions<br />

Access to Funct ons<br />

A r Detector On/Off<br />

Upstream Occlus on Sensor<br />

On/Off<br />

Chang ng the Del very Mode<br />

Yes Yes No No<br />

Yes, w/<br />

code No No No<br />

Yes, w/<br />

code<br />

Yes, w/<br />

code<br />

Yes, w/<br />

code<br />

V ew Only V ew Only V ew Only<br />

V ew Only V ew Only V ew Only<br />

V ew Only V ew Only V ew Only