General<br />
Description<br />
Section 1: General Description<br />
Lock Level Table<br />
This table lists the operations that are accessible in each lock level while<br />
the pump is stopped and running. LL0 permits complete access to all<br />
programming and operating functions. LL1 permits limited control of<br />
pump programming and operations. LL2 permits only minimal control<br />
of pump operations.<br />
*Continuous Delivery Mode = CDM; Intermittent Delivery Mode = IDM<br />
Pump Operations<br />
and Programming<br />
0<br />
Stopped Running<br />
LL0 LL1 LL2<br />
Any Lock<br />
Level<br />
Stop/Start the pump Yes Yes Yes Yes<br />
Reset Reservo r Volume Yes Yes Yes No<br />
Pr me Yes Yes No No<br />
Change the Lock Level<br />
Change Cont nuous Rate<br />
(CDM)*<br />
Clear G ven amount<br />
(CDM & IDM)*<br />
Yes, w/<br />
code<br />
Yes<br />
Yes, w/<br />
code<br />
Up to LL0<br />
value<br />
Yes, w/<br />
code<br />
No<br />
No No<br />
Yes Yes No No<br />
Change Dose Volume (IDM)* Yes No No No<br />
Change Dose Durat on (IDM)* Yes No No No<br />
Change Dose Cycle (IDM)* Yes No No No<br />
Change KVO Rate (IDM)* Yes No No No<br />
Change Dose Starts n Delay<br />
(IDM)*<br />
Biomed Functions<br />
Access to Funct ons<br />
A r Detector On/Off<br />
Upstream Occlus on Sensor<br />
On/Off<br />
Chang ng the Del very Mode<br />
Yes Yes No No<br />
Yes, w/<br />
code No No No<br />
Yes, w/<br />
code<br />
Yes, w/<br />
code<br />
Yes, w/<br />
code<br />
V ew Only V ew Only V ew Only<br />
V ew Only V ew Only V ew Only<br />
V ew Only V ew Only V ew Only