General<br />

Description<br />

Section 1: General Description<br />

Downstream Occlusion Sensor: The pump contains a downstream occlusion<br />

sensor. When a downstream occlusion (between the pump and<br />

patient access site) is detected, an alarm will sound, delivery will stop,<br />

and the display will show “High Pressure.”<br />

Reservoir Volume Alarm: The Reservoir Volume alarm indicates when<br />

the fluid in the fluid container is low or depleted. Each time you change<br />

the fluid container, you may reset the Reservoir Volume to the originally<br />

programmed volume. Then, as medication is delivered, the Reservoir<br />

Volume automatically decreases. When the pump calculates that 5 ml<br />

remain in the fluid container, beeps sound and “ResVol Low” appears on<br />

the main screen. This alarm recurs at every subsequent decrease of 1 ml<br />

until the Reservoir Volume reaches 0 ml, at which point the pump stops<br />

and the Reservoir Volume empty alarm sounds.<br />

The Main Screen<br />

The main screen is the starting point for programming or viewing the<br />

pump’s settings.<br />

If no keys are pressed for a period of time (2 minutes), the display reverts<br />

to the main screen. When the two AA batteries are low, “LowBat”<br />

appears on the main screen.<br />

When running:<br />

Status of pump<br />

Reservoir<br />

Volume<br />

When stopped:<br />

Status of pump<br />

RUN LowBat<br />

ResVol 50.0 ml<br />

STOPPED<br />

Battery Status<br />

Status of<br />

Reservoir<br />

Volume