Section 1: General Description<br />

Å used to put the pump into a low power state when not in use or<br />

back into full power.<br />

Power Jack<br />

You may plug an AC Adapter into the Power jack as an alternate source<br />

of power. The indicator light on the front of the pump will illuminate<br />

when the AC Adapter is in use.<br />

Accessory Jack<br />

This jack is used for accessory cables. See the Instructions for Use supplied<br />

with those accessories.<br />

Air Detector<br />

The Air Detector is on the pump in the area shown in the diagram. If air<br />

is detected in the part of the tubing that passes through the Air Detector,<br />

an alarm sounds and delivery stops. (See Section 5 for Air Detector<br />

specifications.) If an Air Detector is not required, it may be turned off.<br />

(See Section 4, Biomed Functions.)<br />

WARNING: When the Air Detector is turned off, the pump will not<br />

detect air in the fluid path. Periodically inspect the fluid path and<br />

remove any air to prevent air embolism. Air embolism could result in<br />

death or serious injury to the patient.<br />

Cassette<br />

The cassette is the portion of the CADD Medication Cassette Reservoir<br />

or CADD Administration Set that attaches to the bottom of the pump.<br />

The following single-use products are compatible with the CADD-<strong>Legacy®</strong><br />

pump:<br />

• CADD Medication Cassette Reservoir (50 or 100 ml), used with the<br />

CADD Extension Set<br />

• CADD Administration Set