Section 1: General Description<br />
Å used to put the pump into a low power state when not in use or<br />
back into full power.<br />
Power Jack<br />
You may plug an AC Adapter into the Power jack as an alternate source<br />
of power. The indicator light on the front of the pump will illuminate<br />
when the AC Adapter is in use.<br />
Accessory Jack<br />
This jack is used for accessory cables. See the Instructions for Use supplied<br />
with those accessories.<br />
Air Detector<br />
The Air Detector is on the pump in the area shown in the diagram. If air<br />
is detected in the part of the tubing that passes through the Air Detector,<br />
an alarm sounds and delivery stops. (See Section 5 for Air Detector<br />
specifications.) If an Air Detector is not required, it may be turned off.<br />
(See Section 4, Biomed Functions.)<br />
WARNING: When the Air Detector is turned off, the pump will not<br />
detect air in the fluid path. Periodically inspect the fluid path and<br />
remove any air to prevent air embolism. Air embolism could result in<br />
death or serious injury to the patient.<br />
Cassette<br />
The cassette is the portion of the CADD Medication Cassette Reservoir<br />
or CADD Administration Set that attaches to the bottom of the pump.<br />
The following single-use products are compatible with the CADD-<strong>Legacy®</strong><br />
pump:<br />
• CADD Medication Cassette Reservoir (50 or 100 ml), used with the<br />
CADD Extension Set<br />
• CADD Administration Set