Electrosurgical Power Meter - MTK Peter Kron GmbH

Electrosurgical Power Meter - MTK Peter Kron GmbH Electrosurgical Power Meter - MTK Peter Kron GmbH

08.03.2013 Views

EPM 3 Electrosurgical Power Meter The EPM 3 is the ideal choice for all measurement tasks in manufacturing and maintenance of electrosurgical units. OUTPUT POWER AND CURRENT METER FOR ELECTROSURGICAL UNITS Biometric RMS Converter (20mA...2200mA) with internal overload protection Rugged plastic case – lightweight and beautifull Automatic adjustment of the current converter every 30 s Available with calibration certificate according to PTB standards and guidelines of TÜV product service, Munich Large range of load resistors: 10 , 25 ...3175 (128 values) Additional fixed 200 resistor for tests according to EN 60601-2-2 Art. 19.101 (5 k and 10 k optional) Large LC Display with CFL backlight RS 232 Interface Software for Windows PCs with database capability to record data according to manufacturer's instructions Cable sets for diverse electrosurgical units are available upon request Supply Voltage: 100...250 V-AC, 45...60 Hz mtk Peter Kron GmbH Mess-Systeme für Medizintechnik Hard & Software Zossener Straße 55-58, D-10961 Berlin Tel.+49 (0)30 / 69 81 88-40 Fax. +49 (0)30 69 81 88-49 Freeline 0800 040 50 30 www.mtk-biomed.de info@mtk-biomed.de

EPM 3<br />

<strong>Electrosurgical</strong> <strong>Power</strong> <strong>Meter</strong><br />

The EPM 3 is the<br />

ideal choice for<br />

all measurement<br />

tasks in<br />

manufacturing<br />

and maintenance of<br />

electrosurgical units.<br />

OUTPUT POWER AND CURRENT METER FOR<br />

ELECTROSURGICAL UNITS<br />

Biometric RMS Converter (20mA...2200mA) with internal overload<br />

protection<br />

Rugged plastic case – lightweight and beautifull<br />

Automatic adjustment of the current converter every 30 s<br />

Available with calibration certificate according to PTB standards<br />

and guidelines of TÜV product service, Munich<br />

Large range of load resistors: 10 , 25 ...3175 (128 values)<br />

Additional fixed 200 resistor for tests according to EN 60601-2-2<br />

Art. 19.101 (5 k and 10 k optional)<br />

Large LC Display with CFL backlight<br />

RS 232 Interface<br />

Software for Windows PCs with database capability to record data<br />

according to manufacturer's instructions<br />

Cable sets for diverse electrosurgical units are available upon<br />

request<br />

Supply Voltage: 100...250 V-AC, 45...60 Hz<br />

mtk <strong>Peter</strong> <strong>Kron</strong> <strong>GmbH</strong> Mess-Systeme für Medizintechnik Hard & Software<br />

Zossener Straße 55-58, D-10961 Berlin Tel.+49 (0)30 / 69 81 88-40 Fax. +49 (0)30 69 81 88-49<br />

Freeline 0800 040 50 30 www.mtk-biomed.de info@mtk-biomed.de


Specifications*<br />

<strong>Power</strong> supply 100…200 V AC<br />

45…60 Hz<br />

50 VA<br />

safety class 1<br />

Dimensions 443 x298 x 180 mm³<br />

Weight ca. 7,3 Kg<br />

Temperature,<br />

operating +10°C…+40°C<br />

Temperature, storage -20°C…+65°C not condensing<br />

Range 20 mA…2200 mA<br />

Frequency response 100 kHz…5 MHz<br />

(-0,5dB)<br />

Accuracy 20 min after power up: +/-2 %, fot currents < 75 mA: +/-5 mA<br />

Load resistors Accuracy: +/-5%<br />

max. power 10 Ω 135 W<br />

25 Ω 135 W<br />

50 Ω 135 W<br />

100 Ω 270 W<br />

200 Ω 270 W<br />

400 Ω 400 W<br />

800 Ω 270 W<br />

1600 Ω 270 W<br />

edditional fixed resistor<br />

200 Ω 250 W<br />

Serial interface V24/RS 232 C<br />

8 Data bits, 0 Parity,1 Stop bit<br />

Pin assignment<br />

Pin Signal<br />

1 NC<br />

2 TXD<br />

3 RXD<br />

4 NC<br />

5 GND<br />

6,7,8 connected<br />

9 NC<br />

*Specifications are subject to change without notice<br />

9-pol. Sub-D plug<br />

mtk <strong>Peter</strong> <strong>Kron</strong> <strong>GmbH</strong> Mess-Systeme für Medizintechnik Hard & Software<br />

Zossener Straße 55-58, D-10961 Berlin Tel.+49 (0)30 / 69 81 88-40 Fax. +49 (0)30 69 81 88-49<br />

Freeline 0800 040 50 30 www.mtk-biomed.de info@mtk-biomed.de