01.03.2013 Views

TOC Analysis Moving from Laboratory Testing to On-line Analysis

TOC Analysis Moving from Laboratory Testing to On-line Analysis

TOC Analysis Moving from Laboratory Testing to On-line Analysis

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

combined with a ‘main’ and ‘standby’ UV lamp arrangement, so that on detecting the ‘main’ UV lamp<br />

failure, the analyser can au<strong>to</strong>matically switch <strong>to</strong> the ‘standby’ UV lamp and continue <strong>to</strong> moni<strong>to</strong>r on-<strong>line</strong><br />

<strong>TOC</strong> whilst raising an alarm so that the maintenance team can replace the failed UV lamp in a timely<br />

manner.<br />

VERIFYING ON-LINE <strong>TOC</strong> ANALYSERS<br />

Unlike the membrane-based analysers used for labora<strong>to</strong>ry <strong>TOC</strong> analysis the on-<strong>line</strong> <strong>TOC</strong> analyser must<br />

be robust, have a stable calibration and be capable of consistent, repeatable analysis under varying<br />

sample temperatures, pressure and flow-rates.<br />

Figure 4: <strong>TOC</strong> analysers use temperature-compensated conductivity measurement<br />

<strong>TOC</strong> analysers are basically temperature-compensated differential conductivity instruments using UV light<br />

<strong>to</strong> oxidise the organic material <strong>to</strong> CO2 and utilising before-and-after-conductivity measurements as a<br />

surrogate measurement for <strong>TOC</strong>. They must be able <strong>to</strong> accurately moni<strong>to</strong>r the temperature of the sample<br />

as it is analysed <strong>to</strong> ensure that the final conductivity measurement is compensated for temperature<br />

change, otherwise the small change in conductivity caused by the CO2 evolved <strong>from</strong> the oxidised <strong>TOC</strong> will<br />

be swamped by the change in conductivity due <strong>to</strong> the temperature change in the sample.<br />

Hence it is important <strong>to</strong> verify:<br />

a) the calibration of conductivity meter used <strong>to</strong> measure <strong>TOC</strong><br />

b) the calibration of the temperature compensation sensor<br />

c) the <strong>TOC</strong> meter calibration against USP-traceable <strong>TOC</strong> standards<br />

d) system suitability<br />

e) the UV lamps<br />

a) Verifying the calibration of the conductivity meter used <strong>to</strong> measure <strong>TOC</strong><br />

Guidance was given in USP30 regarding the calibration verification conductivity meters. This<br />

was split in<strong>to</strong> two parts: meter calibration and cell constant. Hach Ultra provides an SOP for carrying<br />

these tests out on the Anatel PAT700 <strong>TOC</strong> analyser: SOP 700-4.<br />

Meter calibration verification is made by disconnecting the measurement cell and replacing it with a<br />

known-value NIST-traceable resis<strong>to</strong>r. Cell constant is verified by running a certified conductivity<br />

sample through the measurement cell.<br />

<strong>TOC</strong> <strong>Analysis</strong> – <strong>Moving</strong> <strong>from</strong> labora<strong>to</strong>ry <strong>to</strong> on-<strong>line</strong> moni<strong>to</strong>ring.<br />

March 2008 by Hach Ultra<br />

4 of 7

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!