Mastercycler® ep realplex<br />
Technical specifications<br />
Optical module<br />
Excitation source 96 LEDs (470 nm)<br />
Emission filters 520/550 nm (realplex2 )<br />
520/550/580/605 nm (realplex4 )<br />
Detector 1 photomultiplier tube (realplex2 )<br />
2 photomultiplier tubes (realplex4 )<br />
Dynamic range 8 orders of magnitude from starting copy number<br />
Sensitivity<br />
Thermomodule<br />
≤50 fmol fluorescein<br />
Sample capacity 96 x 0.2 mL PCR tubes or one 96-well PCR plate<br />
(unskirted, semi-skirted, skirted—as per SBS format)<br />
Temperature control range of block 4–99 °C<br />
Degree range of gradient, maximum 1–20 °C (aluminum block)<br />
1–24 °C (silver block)<br />
Temperature control range of gradient 30–99 °C<br />
Temperature of lid 105 °C<br />
Block homogeneity 20–72 °C ±0.3 °C,<br />
90 °C ±0.4 °C<br />
Control accuracy ±0.2 °C<br />
Heating speed* approx. 4 °C/s (aluminum block)<br />
approx. 6 °C/s (silver block)<br />
Cooling speed*<br />
Complete system<br />
approx. 3 °C/s (aluminum block)<br />
approx. 4.5 °C/s (silver block)<br />
Dimensions (W x D x H) 10.24 in x 16.14 in x 15.59 in/26 cm x 41 cm x 39.6 cm<br />
Total weight 52.91 lb/24 kg<br />
Voltage requirements 115 V, 50 Hz–60 Hz<br />
Power consumption 800 W<br />
* Measured at block<br />
Ordering information<br />
Description Catalog No.<br />
Mastercycler ep realplex<br />
Mastercycler ep realplex2 , with aluminum block and two emission filters Please Inquire<br />
Mastercycler ep realplex2 S, with silver block and two emission filters Please Inquire<br />
Mastercycler ep realplex4 , with aluminum block and four emission filters Please Inquire<br />
Mastercycler ep realplex4 S, with silver block and four emission filters Please Inquire<br />
System accessories<br />
Real-time PCR consumables, see page 206 or visit www.eppendorf.ca<br />
Service<br />
Performance plans, see page 214 for details<br />
Please contact <strong>Eppendorf</strong> <strong>Canada</strong> at 800-263-8715 to receive a quotation for the Mastercycler ep realplex system.<br />
Sample Handling 199<br />
In the U.S.: <strong>Eppendorf</strong> North America 800-645-3050 · In <strong>Canada</strong>: <strong>Eppendorf</strong> <strong>Canada</strong> Ltd. 800-263-8715<br />
E-mail: info@eppendorf.com · www.eppendorf.com<br />