DRAGER XL & EVITA VENTILATORS:<br />
SETUP:<br />
see picture below for assistance<br />
• Keep humidifying ventilation circuit intact<br />
• Attach bacterial/viral filters to both inspiratory and expiratory arms of ventilator<br />
• Attach short arm of blue tubing to filter on inspiratory side of ventilator and other end to<br />
humidification bath<br />
• Attach long arm of blue tubing from humidification bath to <strong>CPAP</strong> mask<br />
• Assemble “blue” wire of the humidification bath holder, to blue tubing<br />
• Attach white tubing to filter on expiratory side of ventilator, attach “yellow” wire of the<br />
humidification bath holder to white tubing<br />
• Attach “Y” piece to mask<br />
• When ventilator is in “Stand By” select “Mask” and confirm. If using the EVITA ventilator<br />
(not all EVITA’s can do NIV, check for sticker to indicate) the screen will change from blue<br />
to green and the top right corner will state that the ventilator is in Non Invasive mode. If<br />
using the XL ventilator touch the “Tube/mask” tab and select “ Mask” the words “Mask<br />
Ventilation” should appear in the top right corner in orange<br />
• Check ventilator as per invasive ventilation<br />
• Touch the PEEP dials until desired parameter is reached and confirm<br />
• Attach mask to patient<br />
Setup of <strong>CPAP</strong> on Drager XL ventilator<br />
(Same set up for Drager ventilators with NIV function)<br />
©ROYAL PRINCE ALFRED HOSPITAL INTENSIVE CARE SERVICE